DSI-LVDT
TECHNICAL REFERENCE MANUAL
Figure 6: LVDT adapter output sensitivity to input signal phase lead/lag , 1:1 sensor
Two examples are shown for sensors with sensitivity = 0,5. Output signal level is
1 2 of the Exc. Signal. Sensor
output phase lag is 45⁰. LVDT adapter output is ≈0,35 (≈1Vexc ∙ sensitivity ∙ 0,70) according to Figure 2.
When phase compensation matches phase lead / lag of sensor output LVDT adapter output will be at its
maximum = 0,5 (=1Vexc ∙ sensitivity ∙ 1,00).
DSI-LVDT V20-1