Dewert User Manual 2000XL Series– Updated 05.2021/1 MDR Page 3 of 19
Table operation and patient positioning on the table may only be performed by professionally trained persons
who have been instructed in its use or who, through experience with other similar medical devices, have
knowledge of its proper use, taking into account possible hazards.
Equipped with the options of movability (not wheel-lifting/fixing mechanism), side guards and push handle, the
intended purpose of the table is extended and also provides for the ideal positioning of patients for the
purpose of transport to pre-treatment or post-treatment locations. Patient positioning during the recovery
phase after a medical procedure is also permitted under supervision.
Otherwise, the device must only be moved within the room for cleaning or patient access.
This device has been developed exclusively for use indoors and in normal ambient conditions and can be
used in the following areas: laboratories, medical practices, examination and treatment rooms, hospitals,
clinics, physiotherapy practices, occupational therapy centres and doctors’ surgeries.
This table is not classed as surgical furniture and must, therefore, not be used for surgical purposes.
The expected service life is 10 years or 100,000 drive cycles (double stroke = 1x up and down).
1.5 Information on setup and use
When packed, the device may be exposed to the following environmental conditions for approx. 3 months:
Transport/Storage temperature: -20° to +50°C
Operating temperature: +10° to +40°C
Relative humidity: 30% to 75%
Air pressure: 800hPa to 1060hPa
When transporting the device in a vehicle, it must be secured properly against moving. To do so, lock the
castors (optional equipment) and ensure further safety measures.
When setting up the device, do not lift it at the head part, as this may damage the head part and/or the
release mechanism.
Hold and lift the table on the left and right-hand side of the underframe. The table must stand securely on its
feet or castors on a level, flat and solid surface. Before use, activate the brakes on the castors and make sure
they are working properly.
When transporting the table, take hold of the underframe, NOT the upper frame.
1.6 Commissioning
The device is ready for use upon delivery. Remove the power cable from the film packaging on the
underframe of the table and connect it properly to a permanently installed mains socket. When routing the
power cable, make sure the cable cannot be crushed, rolled over or otherwise damaged.
Lock the castors (optional) or the wheel system (optional). As the operator, carry out a thorough and precise
function check once the device has been set up. Prior to commissioning, clean the device and remove any
contamination from transport. Make sure that no connecting cables from the hand or foot switch to the motor
are trapped in the mechanism and thus damaged.
Operation in potentially explosive atmospheres is not permitted.
1.7 Safety notices
This table may only be used for its intended purpose. Any other use is strictly prohibited and possibly
dangerous. The manufacturer cannot be held responsible for damage caused by improper use. Patients may
only be positioned in preparation for treatment/examination by professionally trained persons.
Please note:
This table is not classed as surgical furniture.
Prior to and when adjusting the height of the table, make sure that no persons or objects are located in the
adjustment range of the table and that nobody has their hands on the underframe
.
The following basically applies: Never reach into or under the frame of the table when adjusting the height.
Height adjustment can result in injury if the user does not pay due care and attention. Therefore, take great
care when performing this procedure.
When adjusting the upholstery parts, make sure that no persons or objects are in the adjustment range. Make
sure that no persons reach under the upholstery part or lean on the underframe.