Dewert User Manual 2000XL Series– Updated 05.2021/1 MDR                                                                Page 3 of 19 

Table operation and patient positioning on the table may only be performed by professionally trained persons 
who have been instructed in its use or who, through experience with other similar medical devices, have 
knowledge of its proper use, taking into account possible hazards. 
Equipped with the options of movability (not wheel-lifting/fixing mechanism), side guards and push handle, the 
intended purpose of the table is extended and also provides for the ideal positioning of patients for the 
purpose of transport to pre-treatment or post-treatment locations. Patient positioning during the recovery 
phase after a medical procedure is also permitted under supervision. 
Otherwise, the device must only be moved within the room for cleaning or patient access.  
This device has been developed exclusively for use indoors and in normal ambient conditions and can be 
used in the following areas: laboratories, medical practices, examination and treatment rooms, hospitals, 
clinics, physiotherapy practices, occupational therapy centres and doctors’ surgeries.  
This table is not classed as surgical furniture and must, therefore, not be used for surgical purposes. 
The expected service life is 10 years or 100,000 drive cycles (double stroke = 1x up and down). 

 
 

1.5  Information on setup and use 

When packed, the device may be exposed to the following environmental conditions for approx. 3 months:   
 
Transport/Storage temperature: -20° to +50°C  
Operating temperature: +10° to +40°C 
Relative humidity: 30% to 75% 
Air pressure: 800hPa to 1060hPa 
       
When transporting the device in a vehicle, it must be secured properly against moving. To do so, lock the 
castors (optional equipment) and ensure further safety measures. 
 
When setting up the device, do not lift it at the head part, as this may damage the head part and/or the 
release mechanism. 
Hold and lift the table on the left and right-hand side of the underframe. The table must stand securely on its 
feet or castors on a level, flat and solid surface. Before use, activate the brakes on the castors and make sure 
they are working properly.  

When transporting the table, take hold of the underframe, NOT the upper frame. 
 

 

1.6  Commissioning 

The device is ready for use upon delivery. Remove the power cable from the film packaging on the 
underframe of the table and connect it properly to a permanently installed mains socket. When routing the 
power cable, make sure the cable cannot be crushed, rolled over or otherwise damaged.  

 

Lock the castors (optional) or the wheel system (optional). As the operator, carry out a thorough and precise 
function check once the device has been set up. Prior to commissioning, clean the device and remove any 
contamination from transport. Make sure that no connecting cables from the hand or foot switch to the motor 
are trapped in the mechanism and thus damaged.  
Operation in potentially explosive atmospheres is not permitted. 
 

 
1.7  Safety notices  

This table may only be used for its intended purpose. Any other use is strictly prohibited and possibly 
dangerous. The manufacturer cannot be held responsible for damage caused by improper use. Patients may 
only be positioned in preparation for treatment/examination by professionally trained persons. 

Please note:

 This table is not classed as surgical furniture. 

 

Prior to and when adjusting the height of the table, make sure that no persons or objects are located in the 
adjustment range of the table and that nobody has their hands on the underframe

.

 

 

The following basically applies: Never reach into or under the frame of the table when adjusting the height. 
Height adjustment can result in injury if the user does not pay due care and attention. Therefore, take great 
care when performing this procedure. 

 

When adjusting the upholstery parts, make sure that no persons or objects are in the adjustment range. Make 
sure that no persons reach under the upholstery part or lean on the underframe. 

 

Summary of Contents for 2000XL Series

Page 1: ...se 3 1 6 Commissioning 3 1 7 Safety notices 3 1 8 Model designation and type labelling 4 1 9 Meaning of the serial number 4 2 Operating manual 5 2 1 Table design 5 2 2 Height adjustment 5 2 3 Head par...

Page 2: ...st important safety information Correct operation of the device is essential for safe operation Therefore please familiarise yourself with the contents of these instructions for use before using the d...

Page 3: ...Hold and lift the table on the left and right hand side of the underframe The table must stand securely on its feet or castors on a level flat and solid surface Before use activate the brakes on the c...

Page 4: ...express permission of the manufacturer When transporting the table take hold of the underframe NOT the upper frame 1 8 Model designation and type labelling The exact model designation depends on the c...

Page 5: ...ke place beforehand see section 7 Alternatively the switch rails can also be led out to the short side of the table This facilitates height adjustment from the short sides of the table The lifting mot...

Page 6: ...end position head down feet up inclined position To lower press down the lying surface and operate the release lever at the same time Once the desired position has been reached let go of the release l...

Page 7: ...ose opening reach under the head part when the table is standing still and press the cushion out from the bottom upwards a slight amount of pressure is required To close it simply insert the cushion i...

Page 8: ...d by tightening the screws The moving parts of the side guard should be re lubricated slightly at regular intervals spray oil e g Ballistol Side guard lowerable for side rail This side guard can be pl...

Page 9: ...sm Comfort twin castors 100 or 125 mm individually lockable Paper roll holder for examination tables Paper roll holder for transport recovery tables Side rails Fastening clamp for side rail Infusion p...

Page 10: ...on equipment kg 75 80 75 80 75 80 75 80 Min max height mm 500 940 520 940 500 940 520 940 Adjustment time motor s 22 22 Head part adjustment range ratchets 0 30 0 30 Head part adjustment range gas sp...

Page 11: ...00 650 800 Head part length mm 750 750 570 750 Weight approx depending on equipment kg 105 105 125 135 125 135 Min max height mm 630 1050 630 1050 570 990 570 990 Head part adjustment range gas spring...

Page 12: ...35XLE 2035XL H Max length mm 1950 1950 Width mm 700 800 700 800 Head part length mm 750 750 Weight approx depending on equipment kg 75 80 75 80 Min max height mm 480 920 500 920 Adjustment time motor...

Page 13: ...es within the meaning of EN 50082 Part 1 and 2 Model 2052XLE 2053XLE is equipped with a second electric motor Both motors are connected to the mains by a power cable 4 3 Technical data for hydraulic s...

Page 14: ...manufactured in such a way that it will operate safely over a very long period of time if used as intended and in a correct manner Depending on the conditions of use place of use and care the expected...

Page 15: ...they cannot be triggered accidentally by patients users or third parties The actuator of this table is equipped with an integrated safety device to protect against unauthorised unintentional operation...

Page 16: ...Packaging Safety notice Pay attention to sharp edges and pointed objects during disposal The packaging materials produced are mainly o Cardboard Paper o Plastic o Wood when delivered on a pallet Pleas...

Page 17: ...th therapy Table operation and patient positioning on the table may only be performed by professionally trained persons who have been instructed in its use or who through experience with other similar...

Page 18: ...damaged All switches and supply lines undamaged Function check With electromotive adjustment Move all motors with the foot switch or the hand switch to both limit positions until automatic switch off...

Page 19: ...ssories Accessories such as belts belt pads belt guides paper roll holders armrests etc undamaged and is proper and secure fixing function possible All necessary toggle screws present Possible movabil...