background image

The “Select Filter” menu is used for adjusting the f ilter mode , of which, there are three modes. The correct and

default setting is “2”. This is a factory authorized adjustment.

3.7  C

ALIBRATION

NOTE: Before calibration, verify the range setting as described in section 3.6.4.

Material Requirements

*

Detcon PN 3270 MicroSafe™ Programming Magnet

*

Detcon PN 6038 Threaded Calibration Adapter 

*

Span gas containing the H2S gas in air. Nitrogen mixtures are not acceptable. H2S gas concentration is recom-

mended at 25 ppm (which is the factory default for ranges 0-50 ppm and 0-100 ppm) or 10 ppm (which is the

factory default for the range 0-20 ppm) at a controlled f low rate of 200 to 500 ml/min. Other concentrations

can be used as long as they fall within allowable levels. See section 3.7.2 for details.

*

Detcon PN 985-241100-321 In-Line Humidifying Tube.

NOTE: Span gas bottles contain 0% humidity and this ultra-low humidity condition will cause inaccurate readings

when used to calibrate a sensor.  To prevent this error, Detcon prescribes the use of an In-Line Humidifying Tube,

which adds the relative humidity to the span gas.  The humidifying tube is not necessary when using a gas generat-

ing calibration device that consists of pumped ambient air and an H2S generating source.

3.7.2  Calibration Procedure - Span

CAUTION:

Verif ication of the correct calibration gas level setting and calibration span gas concentration is

required before “span” calibration. These two numbers must be equal.

Calibration consists of entering the calibration function and following the menu-displayed instructions. The display

will ask for the application of span gas in a specif ic concentration. This concentration is equal to the calibration gas

level setting. The factory default setting for span gas concentration is 10 ppm for the range 0-20 ppm and 25 ppm

for the ranges 0-50 ppm and 0-100 ppm. If a span gas containing the prescribed default concentration is not avail-

able, other concentrations may be used as long as they fall within 10% and 50% of full-scale range. However, any

alternate span gas concentration value must be programmed via the calibration gas level menu before proceeding

with span calibration. Follow the instructions below for span calibration.

a)

Verify the current calibration gas level setting as indicated by the programming status menu. To do this, follow

the instructions in section 3.8 and make note of the setting found in listing number 2. The item appears as

“CalLevel @ xxPPM”.

b)

If the calibration gas level setting is equal to your calibration span gas concentration, proceed to item “f”. If

not, adjust the calibration gas level setting so that it is equal to your calibration span gas concentration, as

instructed in items “c” through “e”.

c)

Enter the programming menu by holding the programming magnet stationary over “PGM 2” for 15 seconds

until the display reads  

“VIEW PROG STATUS”

, then withdraw the magnet. At this point you can scroll

through the programming menu by momentarily waving the programming magnet over “PGM 1” or “PGM 2”.

The menu options are: View Program Status, Set  Cal Level, Set Heater Level, Set Range, and Linearize Sensor.

d)

From the programming menu scroll to the calibration level listing. The menu item appears as: 

“SET CAL

LEVEL”.

Enter the menu by holding the programming magnet stationary over “PGM 1” for 3 seconds until

the display reads  

“CalLevel @ ##PPM”

, then withdraw the magnet. Use the programming magnet to make

an adjustment to “PGM 1” to increase or “PGM 2” to decrease the display reading until the reading is equal to

the desired calibration span gas concentration. Exit to the programming menu by holding the programming

magnet over “PGM1” for 3 seconds.

e)

Exit back to normal operation by holding the programming magnet over “PGM 2” for 3 seconds, or automati-

cally return to normal operation in 30 seconds.

f)

Enter the calibration span menu by holding the programming magnet stationary over “PGM 1” for 3 seconds

The display will read 

“1-SPAN  2-EXIT”

Hold PGM 1 for 3 seconds to proceed with a span or hold PGM 2

for 3 seconds to exit back to normal operation mode. After choosing span you may, within 1 minute, abort the

span procedure by a 3 second hold on PGM 2.

Detcon Model TP-524C Hydrogen Sulfide Sensor

PG.13

Summary of Contents for MicroSafe TP-524C

Page 1: ...n inc Detcon MicroSafe TP 524C Hydrogen Sulfide Sensor This manual covers the following ranges 0 20 PPM 0 50 PPM and 0 100 PPM Operator s Installation Instruction Manual May 09 2012 Document 2297 Version 2 5 ...

Page 2: ...t up 3 7 Calibration 3 8 Status of Programming Software Version Calibration Level Heater Voltage Range and Sensor Life 3 9 Program Features 3 10 Display Contrast Adjust 3 11 Trouble Shooting Guide 3 12 Spare Parts List 3 13 Warranty 3 14 Service Policy 3 15 Software Flow Chart 3 16 Revision Log Detcon Model TP 524C Hydrogen Sulfide Sensor PG 2 ...

Page 3: ...nu driven features via a hand held programming magnet 3 0 1 Sensor Technology The sensor technology is a patented solid state metal oxide semiconductor The sensor consists of two thin films a temperature sensitive heater film and an hydrogen sulfide sensitive sensor film Both films are deposited on a sili con microchip by vacuum deposition The heater film elevates the operating temperature of the ...

Page 4: ...put 3 0 3 Base Connector Board The base connector board is mounted in the explosion proof enclosure and includes the mating connector for the control circuit reverse input and secondary transient suppression input filter and lugless terminals for field wiring 3 0 4 Explosion Proof Enclosure Detcon Model TP 524C Hydrogen Sulfide Sensor PG 4 detcon inc Program Switch 2 FLT CAL MicroSafe H2S Gas Sens...

Page 5: ...ons 3 2 APPLICATION Model TP 524C MicroSafe sensors are designed to detect and monitor hydrogen sulfide gas in ambient air in the range of 0 100 parts per million The sensor can also be programmed to operate in the range 0 20 ppm or 0 50 ppm Minimum sensitivity and scale resolution is 1 ppm Operating temperature range is 40 F to 175 F While the sen sor is capable of operating outside these tempera...

Page 6: ...ace plate The two switches are referred to as PGM 1 and PGM 2 The menu list consists of 3 items which include sub menus as indicated below Note see the last page of this manual for a complete software flow chart 01 Normal Operation a Current Status 02 Calibration Mode a Span 03 Program Menu a View Program Status b Set Calibration Level c Set Heater Level d Set Range e Utility Menu 3 4 1 Normal Ope...

Page 7: ...t X Value is an adjustment for linearization factory setting only Set Filter Mode is a selection of filtering modes factory setting only 3 5 INSTALLATION Optimum performance of ambient air gas sensor devices is directly relative to proper location and installation practice 3 5 1 Field Wiring Table 4 20 mA output Detcon Model TP 524C solid state H2S sensor assemblies require three conductor connect...

Page 8: ...lity engi neers with experience in similar processes Ventilation Normal ventilation or prevailing wind conditions can dictate efficient location of gas sensors in a manner where the migration of gas clouds is quickly detected Personnel Exposure The undetected migration of gas clouds should not be allowed to approach concentrated per sonnel areas such as control rooms maintenance or warehouse build...

Page 9: ...g to the sensor base connector board in accordance with the detail shown in Figure 3 d Replace the plug in transmitter circuit and replace the junction box cover 3 5 6 Remote Mounting Applications Some sensor mounting applications require that the gas sensor head be remotely mounted away from the sensor transmitter This is usually true in instances where the gas sensor head must be mounted in a lo...

Page 10: ...ion 3 6 2 then proceed with the initial operation tests below section 3 6 1 Note 3 The 4 20mA output signal will be held at 4mA for the first two minutes after power up 3 6 1 Initial Operational Tests Initial Operational Tests After a warm up period at least 15 minutes has been allowed for the sensor should be checked to verify sensitivity to H2S gas Material Requirements Detcon PN 6038 Threaded C...

Page 11: ...s or automati cally return to the programming menu in 30 seconds h Exit back to normal operation by holding the programming magnet over PGM 2 for 3 seconds or automati cally return to normal operation in 30 seconds i Replace the junction box cover on the remote sensor enclosure Heater voltage adjustment is complete 3 6 3 Programming Magnet Operating Instructions Operator interface to MicroSafe gas...

Page 12: ... over PGM1 for 3 seconds or automati cally return to the programming menu in 30 seconds e Exit back to normal operation by holding the programming magnet over PGM 2 for 3 seconds or automati cally return to normal operation in 30 seconds NOTE When switching between ranges remember to readjust your Cal Level setting if necessary see section 3 7 2 3 6 5 Setting Heater Voltage The Detcon TP524C H2S s...

Page 13: ... and 50 of full scale range However any alternate span gas concentration value must be programmed via the calibration gas level menu before proceeding with span calibration Follow the instructions below for span calibration a Verify the current calibration gas level setting as indicated by the programming status menu To do this follow the instructions in section 3 8 and make note of the setting fo...

Page 14: ...ignal reports a 0 mA value to indicate there is a Fault condition that needs to be corrected 3 7 3 Additional Notes 1 Upon entering the calibration menu the 4 20 mA signal drops to 2 mA and is held at this level until the pro gram returns to normal operation 2 If during calibration the sensor circuitry is unable to attain the proper span adjustment the sensor will enter into one of the calibration...

Page 15: ...ange 20 50 or 100 ppm it will cause the display to flash 20 50 100 PPM H2S on and off The 4 20 mA signal will report a 20 mA output Zero Fault If the sensor does not clear to 10 of range in 5 minutes after completing a span calibration the sensor will enter into a calibration fault mode which will cause the display to alternate between ZERO FAULT and the sen sors current reading The Fault LED will...

Page 16: ...st remove the enclosure cover and use a jewelers screwdriver to turn the contrast adjust screw located beneath the metallic face plate The adjustment location is marked CONTRAST See figure 6 for location 3 11 TROUBLE SHOOTING GUIDE SENSOR FAULTS Open Heater Verify resistance between black and white wires using ohm meter with the sensor unplugged and at room tempera ture Heater Film s normal readin...

Page 17: ...emperature 2 If poor contrast adjust contrast pot accordingly Nothing Displayed Transmitter not Responding 1 Verify condulet has no accumulated water or abnormal corrosion 2 Verify required DC power is applied to correct terminals 3 Swap with a known good transmitter to determine if transmitter is faulty Bad 4 20 mA output or RS485 Output 1 Check that wiring is connected to correct terminal output...

Page 18: ...hipping point is FOB the Detcon factory Net payment is due within 30 days of invoice Detcon Inc reserves the right to refund the original purchase price in lieu of sensor replacement 3 14 SERVICE POLICY Detcon Inc as manufacturer warrants under intended normal use each new MicroSafe plug in control circuit to be free from defects in material and workmanship for a period of two years from the date ...

Page 19: ...2 3 SET CAL LEVEL PGM2 3 PGM2 3 SENSOR LIFE PGM1 3 PGM2 30 SET X VALUE UTILITY MENU CAL LEVEL PPM HEATER VDC PGM1 M PGM2 M PGM1 3 INC DEC PGM1 3 PGM2 M PGM2 3 HEATER VDC SET HEATER VOLTS PGM1 M PGM2 M PGM1 3 INC DEC PGM1 3 PGM2 M PGM2 3 RANGE 0 PPM SET RANGE RANGE 0 PPM PGM1 3 PGM2 M PGM2 3 SET FILTER PGM1 3 PGM2 M PGM2 3 PGM1 M PGM2 M PGM1 3 INC DEC X PGM1 M PGM2 M PGM1 3 INC DEC FILTER www detco...

Reviews: