10. CHANGING HEPA/ULPA FILTERS 

Disconnect Design-Aire from the power source prior to servicing your unit. Remove the #8-32 Philips tooth lock 
screw and the six catches (4 on 2x2 Design-Aire) holding the filter housing to the HEPA/ULPA filter. Once the 
screws and catches are removed simply lift the filter housing from the HEPA/ULPA filter. Remove your new 
filter from its packaging and visually inspect it for sign of damage. Carefully line up the filter housing with the 
new filter and place on filter. Re-attach the catches and screws. 

 

           Figure 10.1 

 

11. ROOM-SIDE-REPLACEMENT FILTER 

The Design-Aire Room-Side-Replacement filter features a 6.88” anodized aluminum filter housing. A unique 
retaining strip system allows for quick easy filter changes without the use of tools, the gel sealant ensures a 
leak-free uniform seal. Room-Side-Replacement filters are available in HEPA or ULPA filter configurations. 

ROOM-SIDE-REPLACEMENT FILTER INSTALLATION PRECAUTIONS: 

 

Filter installation requires two people 

 

All unit frames must be secured to the “T” bar grid before filters may be installed. 

 

During installation DO NOT exert force on the grill screen or damage will result 

 

Short filter support strips must be installed first on all units except 2’ x 2’. 

 

 

Summary of Contents for DESIGN-AIRE DA2-1HSL-RSR-EC

Page 1: ...DESIGN AIRE SL RSR EC Installation and Operations Manual ...

Page 2: ...GN AIRE SL RSR EC Installation and Operations Manual Revision 28 Jan 16 Page 2 of 15 86 Harry Douglas Dr Ottawa ON K2C 2C7 T 613 831 8318 TF 1 888 636 8609 DESIGN FILTRATION MICROZONE INC www dfmzgroup com ...

Page 3: ...115V MODELS DA2 1HSL RSR EC DA2 1USL RSR EC DA3 1HSL RSR EC DA3 1USL RSR EC DA4 1HSL RSR EC DA4 1USL RSR EC TABLE OF CONTENTS MODELS COVERED 3 1 SHIPPING AND UN CRATING 4 2 FEATURES 5 3 FILTER HOUSING 5 4 VARIABLE SPEED CONTROL 6 5 DUPLEX OUTLET 6 6 POWER CORD 7 7 PRE FILTER 7 8 PRE FILTER MAINTENANCE 7 9 HEPA FILTER 8 10 CHANGING HEPA ULPA FILTERS 9 11 ROOM SIDE REPLACEMENT FILTER 9 12 MOTOR ELEC...

Page 4: ... the plug of your unit if it will not fit in your duplex outlet Have a proper outlet installed by a qualified electrician Use only specified replacement parts recommended by Design Filtration Microzone Use this equipment only in the manner intended by Design Filtration Microzone 1 SHIPPING AND UN CRATING Design Filtration Microzone assembles and factory tests each unit prior to shipping On occasio...

Page 5: ...2 2 3 FILTER HOUSING The filter pre filter housings are manufactured with light weight 16 gauge aluminum with easily removable catch hooks for trouble free filter changes The housing features a screwed in place aluminum screen for the pre filter placement allowing quick and easy pre filter change outs All housings are sealed air tight with RTV silicone The standard variable speed control and duple...

Page 6: ... the recommended airflow range The variable speed control knob is located on the side panel of the pre filter housing opposite the duplex outlet If your Design Aire is equipped with the optional speed control cover you will require a 2 Philips screwdriver to remove the cover Figure 4 1 5 DUPLEX OUTLET Design Aire comes complete with a duplex outlet located on the pre filter housing on the back of ...

Page 7: ...sign Aire is equipped with disposable pre filters which are manufactured from bonded synthetic media set into a moisture resistant beverage board Pleated pre filters supplied with your Design Aire carry an efficiency rating of MERV 7 Minimum Efficiency Reporting Value to change the pre filters simply lift it off the pre filter housing and replace dirty pre filter with new pre filter 8 PRE FILTER M...

Page 8: ... laid glass media is set into the frame with a urethane seal Efficiency is minimum 99 99 at 0 3 micron particle size and larger HEPA filters cannot be cleaned and must be replaced when necessary Your Design Aire may be equipped with an optional ULPA filter which is rated minimum 99 9995 efficient at 0 12 microns and larger ULPA filters cannot be cleaned and must be replaced when necessary Your ind...

Page 9: ...n of damage Carefully line up the filter housing with the new filter and place on filter Re attach the catches and screws Figure 10 1 11 ROOM SIDE REPLACEMENT FILTER The Design Aire Room Side Replacement filter features a 6 88 anodized aluminum filter housing A unique retaining strip system allows for quick easy filter changes without the use of tools the gel sealant ensures a leak free uniform se...

Page 10: ...he ridge on the filter frame is visible from below 12 MOTOR ELECTRICAL Design Aire uses 230 volt energy efficient backward curved impellers to provide excellent airflow distribution quieter noise levels and low heat loads No maintenance is required for the motors although a periodic inspection for dirt build up is recommended Dirt on your motor can cause vibrations and shorten the life cycle of th...

Page 11: ...llation and Operations Manual Revision 28 Jan 16 Page 11 of 15 86 Harry Douglas Dr Ottawa ON K2C 2C7 T 613 831 8318 TF 1 888 636 8609 DESIGN FILTRATION MICROZONE INC www dfmzgroup com 13 WIRING DIAGRAM Figure 12 1 230 Volt Model ...

Page 12: ... EC Installation and Operations Manual Revision 28 Jan 16 Page 12 of 15 86 Harry Douglas Dr Ottawa ON K2C 2C7 T 613 831 8318 TF 1 888 636 8609 DESIGN FILTRATION MICROZONE INC www dfmzgroup com Figure 12 2 115 Volt Model ...

Page 13: ...R EC 23 62 x 47 62 x 11 25 230V 0 25A 60Hz 46lbs 49lbs 12 x 48 x 25 Model Dimensions Electrical Weight Shipping Weight Shipping Dimensions DA2 1HSL RSR EC 23 62 x 23 62 x 11 25 115V 0 8A 50 60Hz 31lbs 33lbs 12 x 24 x 25 DA3 1HSL RSR EC 23 62 x 35 62 x 11 25 115V 1 6A 50 60Hz 42lbs 46lbs 12 x 36 x 25 DA4 1HSL RSR EC 23 62 x 47 62 x 11 25 115V 1 6A 50 60Hz 46lbs 49lbs 12 x 48 x 25 15 AIRFLOW Model S...

Page 14: ...reaker tripped power outage has occurred Reset breaker or replace fuse Do not increase fuse capacity If circuit is overloaded contact qualified electrician Call local electric company for power outage Motor is overheated FFY motor will stop if overheated Motor will automatically restart after cool down period Variable speed control is in off position See page 6 for operation of variable speed cont...

Page 15: ...ation Microzone will not accept charges for removal or re installation Design Filtration Microzone will not accept any returned items without prior written consent Design Filtration Microzone is not responsible for repair or replacement by misuse or abuse by incorrect operating procedures of its equipment Pre filters and carbon filters carry no warranty against loading Any modifications or changes...