Wrist Strap Test:

When power is supplied to the 50513, the “STRAP”

indicator will illuminate red until a functional single wire wrist

strap is worn and plugged into the “WRIST STRAP

MONITOR” jack. Plug a single wire wrist strap into the

connector marked “WRIST STRAP OPERATOR” place the

strap securely around either wrist. the “STRAP” indication

will be green if the total capacitance of operator and wrist

strap is above approximately 50 picofarads and the series

resistance of the wrist strap cord is less than approximately

2 megohms. Should the capacitance or resistance change

above or below these thresholds, the “STRAP” indication

will illuminate red and the alarm will sound for

approximately 3 seconds.

A wrist strap, plugged into the “WRIST STRAP/GUEST”

banana jack, is connected directly to earth ground and is

not continuously monitored.

Because of the inherent nature of single-wire wrist strap

monitoring based on the body capacitance principle, it may

sometimes be necessary to adjust an individual unit to a

particular operator’s body capacitance. Adjust the unit by

removing the top cover of the unit and turning calibration

potentiometer R7 until the “STRAP” indicator just turns red

while the operator is wearing the wrist strap. Then, slowly

turn the potentiometer in the opposite direction until the

“STRAP” indicator just returns to green. Verify that

removing the wrist strap will cause the indicator to turn red

and that the alarm sounds for approximately 3 seconds.

Worksurface Test:

Wear a plugged-in wrist strap to produce a “PASS”

condition for all circuits (all green LEDs are on). Disconnect

the worksurface wire at the rear of the 50513 or at the

bench surface. Observe that the “BENCH” red LED turns

on the the audible alarm is activated for approximately 3

seconds. Reset the alarm circuit by reconnecting the

worksurface wire to the 50513. The green “BENCH” LED

remains on, indicating a bench top OK condition when the

resistance between the worksurface to ground connection

and the worksurface measurement connection from the

50513 is less than 50 megohms.

NOTE:

Special order 50513 units may have special factory

set bench top trip-points and may also include the optional

STANDBY switch. Specially ordered 50513 units may also

be factory set for continuous alarm. The volume of the

audible alarm may be adjusted with potentiometer R37.

Figure 2. Item 50513 Hookup

© 2004 DESCO INDUSTRIES INC.

Employee Owned

Summary of Contents for EMIT 50513

Page 1: ...for a wrist strap by persons visiting the workstation Make the worksurface monitoring connection by attaching two wires to the worksurface Use snap connectors or a CK 1 contact kit with the two contact points positioned approximately 1 or less apart Attach one wire to the pin jack adapter on the back of the unit labeled BENCH and the other wire to system ground Complete the installation by pluggin...

Page 2: ...ual unit to a particular operator s body capacitance Adjust the unit by removing the top cover of the unit and turning calibration potentiometer R7 until the STRAP indicator just turns red while the operator is wearing the wrist strap Then slowly turn the potentiometer in the opposite direction until the STRAP indicator just returns to green Verify that removing the wrist strap will cause the indi...

Page 3: ...D lights The value of resistance that causes the BENCH indicator to change from green to red should be 500 k ohm 20 If the trip point is not within this range adjust R21 for a trip point as close to 500 k ohm as possible If the 50513 was special ordered with a different setpoint test and adjust to this special value so that the trip point is within 20 of the special value 4 Set the decade resistor...

Page 4: ...on MA for a Return Material Authorization RMA and proper shipping instructions and address Include a copy of your original packing slip invoice or other proof of purchase date Any unit under warranty should be shipped prepaid to the Desco EMIT factory Warranty repairs will take approximately two weeks If your unit is out of warranty call Customer Service at 909 627 8178 Chino CA or 781 821 8370 Ca...

Reviews: