TESTING THE AC OUTLET ANALYZER
NOTE: The following procedure should only be done by
someone familiar with voltage hazards. This procedure will
work for 220VAC as long as the neutral and ground are
referenced. 220VAC produced with out of phase 110VAC-
Ground-110VAC will produce a FAIL result.
Isolate the ground plug from the tester by inserting it into a 3
to 2 plug adapter.
A. PASS OUTLET -
Connect a 5 kilohm resistor @ 5%
tolerance between the supply ground
and tester ground. The PASS LED
should remain illuminated until the
resistor is removed. This test confirms
the PASS point for the neutral-to-ground
resistance.
B. FAIL OUTLET -
Connect a 12.5 kilohm resistor @ 5%
tolerance between the supply ground
and tester ground. The FAIL LED
should remain illuminated until the
resistor is removed. This test confirms
the FAIL point for the neutral-to-ground
resistance.
Specifications
Input Voltage:
98132
: 120 VAC ±15%
98131
: 220 VAC ±15%
Current Drain:
< 10mA
Operator Resistance Limits (nominal):
750 kilohms - 35 megohms
Test Voltage, Open Circuit:
27 VDC (nominal)
Long Term Drift:
< 1% / decade
Dimensions:
98132
: 2.0" x 3.0" x 1.3" (5 cm x 8 cm x 3 cm)
98131
: 2.0" x 3.0" x 2.0" (5 cm x 8 cm x 5 cm)
Operating Temperature:
0° to 40°C