Desco 19770 Operation And Maintenance Download Page 2

TB-2024

Page 2 of 4

Inspection

Remove the meter from the carton and
inspect for damage. Each test kit should
include the following:

1 Protective carrying case
1 LCD megohmeter
2 Test leads
2 5-pound electrodes
1 Grounding clip
1 AC adapter/charger
1 Rechargeable battery

Properly store the meter and component
assemblies when not in use. Do not
charge battery unless it is fully
discharged. Doing so will reduce the life
of the battery.

Features

A. Scale: 

1/2 inch high digital LCD

display provides easy to read
resistivity/resistance measurements.
Values are expressed with a mantissa
and exponent power. For example, if
the display reads “2.9 E 9”, 2.9 is the
mantissa, 9 is the exponent power; the
reading is 2.9 x 10

9

ohms.

B. Test Button: 

This button turns on

the power to the meter. The operator
needs to hold the button down for 10-
20 seconds for the unit to make
numerous readings and calculations.
(Testing in accordance with ESD 
S4.1-1997 requires 15 seconds of
electrification; in contrast, most analog
type meters display measurements
instantaneously.) When released, the
measured reading will remain illuminated
for at least 10 seconds.

C. Test Range Voltage Switch:

10 Volts for 1x10

3

- 1x10

11

ohms

100 Volts for 1x10

6

- 1x10

12

ohms

D. Jacks:

2.5mm plugs will fit the

megohmmeter jacks. Banana plugs will
fit the 5 lb. electrodes.

E. AC-Battery Charger Jack:

12 volts DC - 100mA.

F. 9 Volt Battery Compartment.

G. Parallel Electrodes.

Figure 3. Features of the LCD Surface Resistance/Resistivity 
Megohmmeter

Figure 4. If the value of the
measurement is below 1000 ohms on
the 10 volt scale or 1x10

6

on the 100

volt scale an “L” will appear on the LCD
display.

Figure 5. If the value of the
measurement is over 5x10

12

ohms an

“H” will appear on the LCD display.

Figure 6. A “P” will appear on the LCD
display if there is too much electrical
60Hz noise in the area or if a test is
performed on a high open resistance.

Figure 7. When the unit’s batteries
need recharging a “BAL” will appear on
the LCD display as the meter is being
operated. To maximize the life of the
battery do not recharge until a low
battery “BAL” is displayed.

A

B

E

D

G

C

F

1.7x10

5

S

Item #19770

Chino, CA

Canton, MA

DESCO WEST

- 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178 • Fax (909) 627-7449 

DESCO EAST

- 90 Hudson Road, Canton, MA 02021-1407 • (781) 821-8370 • Fax (781) 575-0172 • Web Site: http://www.desco.com

Summary of Contents for 19770

Page 1: ...5 pound electrodes item 50003 and test leads item 50075 can be bought separately The Test Kit displays a measurement per ESD S4 1 1997 proving the proper electrification period of 15 seconds during which numerous readings and calculations are executed NOTE Testing in accordance with ESD S4 1 1997 requires 15 seconds of electrification in contrast most analog type meters display measurements instan...

Page 2: ...he measured reading will remain illuminated for at least 10 seconds C Test Range Voltage Switch 10 Volts for 1x103 1x1011 ohms 100 Volts for 1x106 1x1012 ohms D Jacks 2 5mm plugs will fit the megohmmeter jacks Banana plugs will fit the 5 lb electrodes E AC Battery Charger Jack 12 volts DC 100mA F 9 Volt Battery Compartment G Parallel Electrodes Figure 3 Features of the LCD Surface Resistance Resis...

Page 3: ...re selected Set the meter to the required test voltage determined by the test procedure and the resistance of the material Conductive material 104 ohms or less is usually tested at 10 volts Dissipative material is usually tested at 100 volts If material to be tested is unknown it is recommended to first test at 10 volts C Press and hold the red test button for 10 20 seconds the unit will be making...

Page 4: ... the resistors to the enclosed cords using grounding clips and recording the meter display Keep the cords separated Adjustments to the internal resistance pots can be done quite easily Request this procedure from the Desco Customer Service Department Specifications Ranges 103 1011 ohms 10 Volts 106 1012 ohms 100 Volts Power Supply 9 volt rechargeable Nickel Cadmium battery and AC adapter charger i...

Reviews: