Observe the LEDs for the proper response as indicated. Be sure 

to hold the cord at an insulated point, so that resistance value is 

not affected by the body. Should testing reveal that the Tester is 

not functioning properly, verify that the battery or power supply is 

operating correctly.
Tester calibration can also be verified with the use of discrete 

resistors and two banana-to-alligator cords.

Adjustment

Detailed adjustment instructions follow:
Equipment Required: - Small blade screwdriver or Trimpot 

adjusting tool, manufacturer’s calibration box or a resistor decade 

box or discrete resistors, with appropriate test leads.

PREPARATION OF THE UNIT:

Make sure that a fresh 9 Volt alkaline battery is installed, and 

orient tester so that it corresponds to Figure 15.

RANGE

FAIL LO PASS LO   PASS HI FAIL HI

WRIST STRAP

675K

    825K 

8.5M

11.5M

FOOT WEAR

675K

825K

80M

120M

RESISTANCE

“A”

“B”

“C”

“D”

CALIBRATION OF FOOT WEAR RANGES:

1.  Connect resistance “A” from  table 1 or 2, this time between

 

the test button and the “FOOT PLATE” banana jack on the  

 

left side of the  tester.  For footwear calibration, press and  

 

hold the touch plate.  The red “FAIL LO” LED should be on.

2.  If required, adjust 

FWL

 trimpot to the point where the “FAIL

 

LO” LED illuminates.

3.  Connect resistance “B” and press switch as in step 1.  The

 

green “PASS” LED illuminates.  The 

FWL

 trimpot may

 

require a slight re-adjustment.

4.  Connect resistance “C” (

footwear range

) and press switch as

 

in step 1.  The green “PASS” LED should illuminate.  If

 

required, adjust the 

FWH

 trimpot to the point where the green

 

LED illuminates.

5.  Finally, connect resistance “D” (

footwear range

) and press

 

switch as in step 1.  The red “FAIL HI” LED should illuminate.

 

The 

FWH

 trimpot may require a slight re-adjustment.

6.  Re-check to insure that all footwear ranges are correct.

RETURN UNIT TO SERVICE

.

Unauthorized modifications will void the product warranty. 

Servicing should be performed only at the factory.  See warranty 

section for repair information.

Specifications

Wrist Strap Circuit

FAIL HI 

Factory set at 11.5 Megohms

FAIL LO 

Factory set at 675 Kilohms

Footwear Circuit

FAIL HI 

Factory set at 120 Megohms

FAIL LO 

Factory set at 675 Kilohms

General Characteristics

Power 

9 Volt battery or optional AC power supply                                                      

Operation 

Resistance bridge

Readout 

Three LED’s & audible alarm                                                                                               

Accuracy 

±20%

Weight 

12 pounds

Height 

39 inches 

 

Consider adding the 

19250

 to your list of “test equipment [that] 

shall be selected to make measurements of appropriate properties 

of the technical requirements that are incorporated into the ESD 

program plan” as required by paragraph 6.1.3.1 of ANSI/ESD

S20.20.

Figure 15 . Back of the Combo Tester

CALIBRATION OF WRIST STRAP RANGES

:

1.  Connect resistance “A” from Table 1 OR 2 between the touch

 

plate and the “WRIST CORD” banana jack.  

A banana plug

 

must be inserted in the banana jack to properly test the

 

wrist strap ranges.

  Press and hold the test button.  The red

 

“FAIL LO” LED should illuminate.

2.  If required, adjust 

WSL

 trimpot to the point where the “FAIL

 

LO” LED illuminates.

3.  Connect resistance “B” and press the switch as in Step III.

 

The green “PASS” LED should illuminate.  The 

WSL

 trimpot

 

may require a slight re-adjustment.

4.  Connect resistance “C” (

wrist strap range

) and press the

 

switch as in step 1.  The green “PASS” LED should be on.  If

 

required, adjust the WSH trimpot to the point where the green

 

“PASS” LED illuminates.

5.  Finally, connect resistance “D” (

wrist strap range

) and press

 

the swtich as in step 1.  The “FAIL HI” red LED should be

 

on.  The 

WSH

 trimpot may require re-adjustment.

6.  Re-check to insure that all ranges are correct, then proceed

 

to calibrate footwear ranges.

TABLE 1 - STANDARD RANGES

Summary of Contents for 19250

Page 1: ...ot intended for the test Parallel ground paths may be caused by several different situations For example a The path represented by the person standing with one shoe on the floor and the other shoe on the test apparatus A parallel path may be created by the shoe on the floor Current from the test instrument is then directed down two paths when it was intended to be directed down one The correct pat...

Page 2: ...ct out more than 1 4 from mounting surface The template on page four is actual size 2 Mount the Tester on the screws pulling down to lock it in place Figure 3 Stationary installation of the Combo Tester Operation The Combo Tester can be operated either on battery or AC power The unit comes equipped with a 9 volt alkaline battery For AC operation plug the optional AC adapter into the mini phone jac...

Page 3: ...HI LEDs light and the audible indicator sounds the wearer should check the foot grounding device immediately 8 Repeat steps 3 6 with other foot Free Standing Test Fixture Assembly and Operation Model 19252 The 19252 s rugged steel pedestal tube is powder coated in a non conductive white finish that helps to prevent false readings if contacted by skin or loose smocks 3 Open the compartment and atta...

Page 4: ...st be calibrated individually Desco offers a calibration unit that is specifically designed to simplify the calibration procedure This unit is sold as item 07010 The Calibration Unit comes calibrated to NIST traceable standards For additional information on the 07010 ask for Technical Bulletin TB 2039 750 Kilohm 10 Megohm TEST RANGE WRIST STRAPS The following resistance should give the display sho...

Page 5: ...switch as in step 1 The red FAIL HI LED should illuminate The FWH trimpot may require a slight re adjustment 6 Re check to insure that all footwear ranges are correct RETURN UNIT TO SERVICE Unauthorized modifications will void the product warranty Servicing should be performed only at the factory See warranty section for repair information Specifications Wrist Strap Circuit FAIL HI Factory set at ...

Page 6: ...ns and address Desco will quote repair charges necessary to bring your unit up to factory standards Warranty Exclusions THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUCT WARRANTIES EXPRESSED AND IMPLIED INCLUDING MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WHICH ARE SPECIFICALLY DISCLAIMED The express warranty will not apply to defects or damage due to accidents neglect ...

Page 7: ... 401 Building No 5 JiuGe Business Center Lane 2301 Yishan Rd Minhang District Shanghai China Sales Direct 86 21 6401 6692 Email amall ameya360 com QQ 800077892 Skype ameyasales1 ameyasales2 Customer Service Email service ameya360 com Partnership Tel 86 21 64016692 8333 Email mkt ameya360 com ...

Reviews: