8.  Plug the power adapter into the outlet and connect its  

  plug into the power jack located at the back of the Control 

  Unit.  The green worksurface LEDs should illuminate. If 

  the red worksurface LED(s) illuminate and alarm, inspect 

  the snap fastener and ground cords for proper connection. 

  The operator LEDs will illuminate red when an operator is 

  not connected. The Dual Operator Continuous Monitor is 

  now ready for use.

Installation on Micastat

®

 Laminate

Materials Needed:

2  Flush Mount Ground Laminate Inserts

1  Installation Tool for Flush Mount Insert
1.  Install the Flush Mount Inserts first. It is recommended to 

  install one in the right back corner and the other in the left 

  back corner of the Micastat

®

 laminated bench top.

2.  Once both inserts are installed, cut the ground cord 

  supplied with the monitor into two pieces; make sure that

  the tinned end is long enough to connect from the 

  worksurface terminal on the Satellite Remote to one of the 

  flush mount terminals located under the bench.
3.  The remaining wire can then be used to connect the other 

  flush mount terminal to ground using the supplied ring 

  terminal.  The shrouded molded end can then be cut off.
When complete, one flush mount terminal will connect the 

Micastat

®

 laminate to ground while the other will connect 

it to a satellite remote.  The satellite remote then monitors 

the ground connection between the two inserts, and since 

Micastat

®

 is the only ESD laminate made with a physical 

uniform ground layer, it will monitor the entire Micastat

®

 

surface ground.  The wires will stay under the bench, 

maximizing the usable space on the bench top.

Operation

When the satellite remote unit is installed and connected 

to a grounded ESD protective worksurface, the monitor’s 

green worksurfaces LED’s and red operator LED’s should 

be illuminated. The monitor takes 6 seconds to activate its 

audible alarm circuitry when first powered.

To use the monitor

:

1.  Plug a wrist strap cord, into the jack labeled OPERATOR 

  on the satellite remotes.  This automatically activates the 

  selected operator channel.  The red operator LED should 

  illuminate, and the buzzer should sound.
2.  Remove the charges from your body by making contact 

  with a properly grounded ESD worksurface. Snap the 

  cord to the wristband, which should fit snugly on your 

  wrist.  This should silence the alarm and cause the LEDs 

  to switch from red to green.  If this does not happen, 

  check the wrist cord for continuity or damage.  Examine 

  your wristband and ensure that it has a secure fit.  If you 

  have dry skin, use ESD hand lotion such as Menda 

 

35666

 Reztore™ ESD Hand Lotion.

When leaving the area, the operator may take the wrist cord 

along or unsnap the cord from wristband and attach it to 

the parking snap located on the satellite remote. This will 

de-activate the alarm and allow the operator to leave the 

workstation. If the cord is plugged into the OPERATOR jack, 

the monitor will alarm continuously until the cord is either 

parked or unplugged.

WORKSURFACE CHANNEL

The Dual Operator Continuous Monitor’s worksurface 

monitoring circuitry is sufficiently sensitive to detect 

extremely low current, allowing it to be used with mats having 

a resistance of up to 1 x 10

7

 ohms (10 megohms). If the user 

desires not to monitor the worksurface, the circuit can be 

bypassed with the switch located at the rear of the satellite. 
When the monitor is connected to a static dissipative 

worksurface, the amount of current that flows is a function of 

the total resistance between the monitor and the worksurface 

to ground. When the resistance of the worksurface is below a 

present threshold, the monitor will indicate good. Conversely, 

if the resistance level is high when compared to the monitor’s 

reference, the unit will alarm. This is an integrating resistance 

measuring circuit. Therefore, it is relatively insensitive to 

externally induced electromagnetic fields. The resistance 

threshold is factory set to 10 megohms (1 x 10

7

 ohms). 

Custom worksurface range is available upon request. 

Contact customer service for more details.

Note

: Per ESD S6.1, Grounding paragraph 4.1.1 “Every 

element to be grounded at an ESD protected station shall 

be connected to the same common point ground.” The 

ESD control plan should include a tailoring** statement 

that continuous monitors are used and that consistent with 

manufacturer’s design grounding is via AC adapter cord to 

equipment ground.

Maintenance and Calibration

The Dual Operator Workstation Continuous Monitor is solid 

state and designed to be maintenance free.  It is calibrated 

to NIST traceable standards.  There are no user adjustments 

that can be made.  Because of the impedance sensing 

nature of the test circuit, special equipment is required 

for calibration.  We recommend using the Desco 

98220

 

Continuous Monitor Calibration Unit whenever calibration 

is performed annually or any of the satellite remotes are 

replaced.  The Calibration Unit is a most important product 

which allows the customer to perform NIST traceable 

calibration on continuous monitors and can be used to 

perform compliance verification required by ANSI/ESD 

S20.20 Section 6.2.2.2.  The Desco 

98220

 is designed to be 

used on the shop floor at the workstation, virtually eliminating 

downtime, verifying that the continuous monitor is operating 

within tolerances.

Summary of Contents for 19246

Page 1: ...periodic or pulsed testing and can save a significant amount of money in testing costs and rejected product Periodic testing detects wrist strap failures after ESD susceptible products have been manufactured The costs of dealing with the resulting catastrophic failures or latent defects can be considerable Dual Operator Workstation Continuous Monitors eliminate the need for users to test wrist str...

Page 2: ...example during each 2 2 second pulse cycle of a leading constant resistive monitor electrical current is pulsed for only 0 2 seconds followed by an unmonitored interval of 2 seconds This leaves the user wrist strap unmonitored for over 90 of each cycle Damaging ESD events can easily occur in the portion of the time in between the pulses The off period of 2 seconds equals 2 billion nanoseconds and ...

Page 3: ...ellite Remotes to the Control Unit using the included cables Power Adapter Workstation 2 Workstation 1 Figure 6 Typical setup of the Dual Operator Continuous Monitor and Satellites To Satellite Mat Monitor Cord Mat Ground Cord Worksurface Mat or Screw allows ground cord to be bolted to mat and prevent it from disconnecting Figure 7 Installing the Mat Ground and Mat Monitor Cords 2 85 4 63 5 13 1 1...

Page 4: ...xamine your wristband and ensure that it has a secure fit If you have dry skin use ESD hand lotion such as Menda 35666 Reztore ESD Hand Lotion When leaving the area the operator may take the wrist cord along or unsnap the cord from wristband and attach it to the parking snap located on the satellite remote This will de activate the alarm and allow the operator to leave the workstation If the cord ...

Page 5: ...OREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUCT WARRANTIES EXPRESSED AND IMPLIED INCLUDING MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WHICH ARE SPECIFICALLY DISCLAIMED The express warranty will not apply to defects or damage due to accidents neglect misuse alterations operator error or failure to properly maintain clean or repair products Limit of Liability In no event wil...

Reviews: