Testing

Lab environment should be at 75 °F at ±5 °F (23.9 °C at 

±5 °C), relative humidity at 45 % at ±5 % RH. Observe 

offset voltage for no less than 10 seconds before 

recording.

Specification

Recorded 

value

Balance

25 V

Discharge time

(+1,000 V to +100 V)

≤ 2 s

Discharge time

(-1,000 V to -100 V)

≤ 2 s

Adjustment

Adjustments to the offset voltage could be made by 

using an insulative flat head screw. Balancing pot is 

located on the side of controller, where the white plastic 

cap is covering it.

Specifications

Input voltage and 

frequency

220 VAC, 50/60 Hz

Neutralisation 

(discharge) time at

6" (15 cm), 30 psi

< 2 seconds

Balance at 15 cm,

30 psi

±25 V

Power consumption

2.5 W (standby)

25 W (in-use)

Amperage

< 50 mA (standby)

< 400 mA (in-use)

Fuse

1 A, 5 x 20 mm, fast-acting

(2 each)

Ion emission

AC

Emitter point

Stainless steel

Ozone

< 0.005 ppm

Air hose

ABS plastic, 2 m length

Gas input

30 to 100 psi, Clean Dry Air 

(CDA)

Gas connection

1/4" O.D. tubing with 1/8" 

NPT with adjustable needle 

valve

Gas consumption

3.5 CFM @ 30 psi

Mounting

Metal mounting plate attaches 

to back of console

Enclosure

Powder coated steel and 

stainless steel

Dimensions (console)

5.1" x 6.4" x 7.1"

(130 mm  x 163 mm x 180 

mm)

Weight (console)

11 lbs (5.0 kg)

Certifications

Intertek, CE

Country of origin

United States of America

Adjustment screw

Figure 9. Balance adjustment trimpot

Limited Warranty, Warranty Exclusions, Limit 

of Liability and RMA Request Instructions

See the Desco Europe Warranty - 

DescoEurope.com/Limited-Warranty.aspx

Summary of Contents for 19588

Page 1: ...ors and isolated semiconductors Some examples of process necessary insulators are the PC board itself plastic test stands plastic housing where a PCB may be mounted as well as computer monitor screens and regular cleaning wipes Examples of floating or isolated conductors are loaded PCB mounted in a stand where the pins are not contacting the dissipative workstation Ionisation is not effective on i...

Page 2: ...lter from quick connect fitting Push against the collar with thumbnail While holding collar in push tubing inward slightly then pull tubing out IMPORTANT NOTE User should exercise caution when using any compressed air device Electrical Requirements The Ion Python and Neutralising Hand Gun must be connected to a properly grounded receptacle for the units to operate properly It is recommended that t...

Page 3: ...izer service Routine service is typically required to meet quality compliance verification requirements Under normal conditions the ioniser will attract dirt and dust especially on the emitter electrodes To maintain optimum performance cleaning must be performed on a regular basis The electrodes should be cleaned at least every six months However more frequent cleaning may be required if used in e...

Page 4: ...sembly simply repeat the process in reverse NOTE Make sure the fittings inside the connector are properly aligned with their receptacles before tightening the threaded collar Replacing the Air Filter of the Ion Python Periodic replacement of the air filter is recommended for optimum performance of the ioniser Compress outer sleeving away from the filter housing Examine the filter for any evidence ...

Page 5: ... 50 60 Hz Neutralisation discharge time at 6 15 cm 30 psi 2 seconds Balance at 15 cm 30 psi 25 V Power consumption 2 5 W standby 25 W in use Amperage 50 mA standby 400 mA in use Fuse 1 A 5 x 20 mm fast acting 2 each Ion emission AC Emitter point Stainless steel Ozone 0 005 ppm Air hose ABS plastic 2 m length Gas input 30 to 100 psi Clean Dry Air CDA Gas connection 1 4 O D tubing with 1 8 NPT with ...

Reviews: