end range, in either flexion or extension. The
active end range is the extent to which the
wrist can be flexed or extended without help.
6. Engage gearing mechanism: Pull the black
engagement knob (1) outward and pivot the
position adjustment knob (2) forward
towards the hand. The black engagement
knob will click into place. Note: To fully
engage the gearing mechanism, the position
adjustment knob (2) may need to be rotated
slightly.
7. Ulnar/Radial Deviation: Adjust for
Ulnar/Radial Deviation by using the Allen
wrench, and loosening the two screws (10)
on the Ulnar-Radial plate (9). Adjust to
desired position, and re-tighten the screws.
8. Adjusting the
orthosis position:
Rotate the position
adjustment knob (2)
until the joint’s soft
tissues are stretched
to patient tolerance.
Note: See Patient Protocol Card to determine
the appropriate amount of stress, and
wearing time.
9. Removing orthosis: Disengage the gearing
mechanism by pulling out the engagement
knob (1), and at the same time pivot the
position adjustment knob (2) toward the
back of the unit. The handcuff should move
freely. Unfasten the forearm straps.
10. Changing softgoods: Remove the four
screws from the forearm plate (11) to
remove the forearm cuff. Attach new
forearm cuff. Remove the two plate screws
(10) from the Ulnar-Radial plate to remove
the handcuff. Attach new handcuff.
CARE INSTRUCTIONS
Clean Wire-Foam
™
pads and exterior of hinge with
non-abrasive cleaner. Straps and blue liner may
be machine-washed. Air-dry liner and straps prior
to wearing.
PRECAUTION
Caution should be used if any of these conditions
exist: peripheral vascular disease, healing skin
wounds, severe spasticity, heterotrophic
ossification, and mild osteoporoses.
Manufacturer
DeRoyal
200 DeBusk Lane, Powell, TN 37849 U.S.A.
888.938.7828
865.938.7828
Authorized Representative
DeRoyal Europe, Ltd.
Virginia Road, Kells, County Meath Ireland
353.46.9280100
Part #1-4415
Revision 03 Revised 2/05
©2005 DeRoyal. All Rights Reserved.
8
end range, in either flexion or extension. The
active end range is the extent to which the
wrist can be flexed or extended without help.
6. Engage gearing mechanism: Pull the black
engagement knob (1) outward and pivot the
position adjustment knob (2) forward
towards the hand. The black engagement
knob will click into place. Note: To fully
engage the gearing mechanism, the position
adjustment knob (2) may need to be rotated
slightly.
7. Ulnar/Radial Deviation: Adjust for
Ulnar/Radial Deviation by using the Allen
wrench, and loosening the two screws (10)
on the Ulnar-Radial plate (9). Adjust to
desired position, and re-tighten the screws.
8. Adjusting the
orthosis position:
Rotate the position
adjustment knob (2)
until the joint’s soft
tissues are stretched
to patient tolerance.
Note: See Patient Protocol Card to determine
the appropriate amount of stress, and
wearing time.
9. Removing orthosis: Disengage the gearing
mechanism by pulling out the engagement
knob (1), and at the same time pivot the
position adjustment knob (2) toward the
back of the unit. The handcuff should move
freely. Unfasten the forearm straps.
10. Changing softgoods: Remove the four
screws from the forearm plate (11) to
remove the forearm cuff. Attach new
forearm cuff. Remove the two plate screws
(10) from the Ulnar-Radial plate to remove
the handcuff. Attach new handcuff.
CARE INSTRUCTIONS
Clean Wire-Foam
™
pads and exterior of hinge with
non-abrasive cleaner. Straps and blue liner may
be machine-washed. Air-dry liner and straps prior
to wearing.
PRECAUTION
Caution should be used if any of these conditions
exist: peripheral vascular disease, healing skin
wounds, severe spasticity, heterotrophic
ossification, and mild osteoporoses.
Manufacturer
DeRoyal
200 DeBusk Lane, Powell, TN 37849 U.S.A.
888.938.7828
865.938.7828
Authorized Representative
DeRoyal Europe, Ltd.
Virginia Road, Kells, County Meath Ireland
353.46.9280100
Part #1-4415
Revision 03 Revised 2/05
©2005 DeRoyal. All Rights Reserved.