![DENTSPLY thermaprep plus Instruction Manual Download Page 3](http://html1.mh-extra.com/html/dentsply/thermaprep-plus/thermaprep-plus_instruction-manual_2483563003.webp)
The oven is designed with two precision heating alerts:
1. First size LED button blinks: The targeted temperature of the heat-
ing element was not reached within specified time (time-out). Use
the other heating element to warm the obturator.
This may be re-set by turning the power off and turning the
power on. If the error occurs again the oven should no longer be
used and sent for repair.
2. Second size LED button blinks: The hardware safety circuit has
been activated due to a failure in the heating element, causing
overheating. This may be re-set by turning power off, discon-
nect the power cord, reconnect the power cord, and turn the
power on. If the error occurs again the oven should no longer
be used and sent for repair.
Clean oven after each use. Any gutta-percha remaining on the
holder or in the heating chamber may be effectively removed with
Orange Solvent (generic dental solvent) on a cotton pledget. Press
and hold both buttons at once for 2 seconds to put the oven into
the cleaning mode. This will cause both heaters to heat up to a
medium temperature of 140°F/60°C. The indicator lights around
the buttons will show that the heaters are on (constantly on). A
signal “beep” and flashing indicator lights around the buttons will
warn that the desired temperature has been reached. Disconnect
the main plug and clean the heater chambers with a wire brush, as
directed on the last page of the oven illustration.
To abort the cleaning mode without disconnecting the main plug and
to enter the normal mode again press either button for 2 seconds.
1. To prevent fire or electric shock, do not expose this product to
moisture.
2. Unplug the oven when not in use.
3. Do not remove the cover of the oven.
4. This oven is designed for use in the dental office. Do not use in
any other environment.
5. To reduce the risk of electric shock, do not remove cover; there
are no user-serviceable parts inside.
6. This oven is intended for use in heating gutta-percha coated
obturators only. Do not insert anything other than an obturator
into the oven.
7. Changes or modifications not expressly approved by DENTSPLY
Tulsa Dental Specialties could void the user’s authority to oper-
ate the equipment.
Voltage rating; 115VAC, 50/60 Hertz, range of the supply voltage
+/- 10%, power-save stand-by mode:
<
1.5W normal use: 380W max
Fuse Rating: 4AT 250V
This equipment has been tested and found to comply with the limits for
a Class B digital device, pursuant to part 15 of the FCC rules. These
limits are designed to provide reasonable protection against harmful
interference in a residential installation. This equipment generates, uses
and can radiate radio frequency energy and, if not installed and used
in accordance with the instructions, may cause harmful interference to
radio communications. However, there is no guarantee that interference
will not occur in a particular installation. If this equipment does cause
harmful interference to radio or television reception, which can be de-
termined by turning the equipment off and on, the user is encouraged to
try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and the receiver.
• Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
Caution: Hot Surface
Class II Equipment
Caution: See Directions for Use
Consult Instructions for Use
Do Not Throw Away
Manufacturer’s Information
Technical
Specifications
Explanation of
Marking Symbols
NOTE:
Precision Heating
Alerts
Cleaning Instructions
Warning
Precautions