6
CONTROL DESCRIPTIONS
• Temperature Indicator:
(1) The temperature on the Miniglaze/2 indicates temperature within ±10°C (±18°F). the
indicator has both Fahrenheit and Centigrade graduations.
• Temperature Control Dial:
(2) The temperature control dial has both Fahrenheit and Centigrade graduations. Dial
markings range from 150°C (300°F) to 1093°C (2000°F) in increments of 25° Fahrenheit and
Centigrade.
• Tone/Light Switch:
(3) The tone/light switch provides a choice between a tone with a light, or a light only to indicate
when the set temperature is attained. When the switch is in the LIGHT position, a red light (7) will
be illuminated on the temperature control dial when the muffle reaches a temperature 10°C below
set point. When the switch is in the TONE position the red light is accompanied by a tone.
• Muffle Power Light:
(4) A red light located above the temperature indicator signals when power is being applied
to the muffle. It will blink intermittently during temperature rise due to the automatic temperature
control.
• Temperature Rise Control Dial:
(5) The Miniglaze/2 is designed to provide an approximate straight line
temperature rise. Complete counterclockwise adjustment ("S", Slow setting) will give an
approximate 10°C (18°F) per minute temperature rise. Complete clockwise adjustment ("F", Fast
setting") will give an approximate 100°C (180°F) per minute temperature rise. (
NOTE
: During initial
heat-up, the rate of increase will be slow to prevent overshoot).
• Calibration Screw:
(6) Adjust for changes in thermocouple. (See Calibration).
7
6
5
2
3
4
1
ON
OFF
LIGHT
TONE
CALIBRATION
RATE