Congratulations!

The ProSystem GT complete endodontic system you have decided to put 

into practice is truly revolutionary. 

An instructional DVD is included in your system purchase. Please 

view this DVD, in addition to reading this booklet, to gain a complete 
understanding of the system and the technique. If you would like to speak 
with one of our staff clinicians, please call 1-800-662-1202, and we’ll 
provide one-on-one assistance.

InDIcatIons For Use

For the removal of dentin, shaping and obturating of the root canal.

contraInDIcatIons

None.

WarnIngs

this product contains dry natural rubber.
rotary files are non-sterile products and must be sterilized  

  before use.

to prevent fire or electric shock, do not expose 

  the oven to moisture.

Do not remove the cover of the oven.
Do not use the oven near standing water. Dropping the 

  oven and/or motor into water may cause electric shock

turn off oven and unplug when not in use.
this oven is designed for use in a dental office. Do not use  

  in any other environment.

PrecaUtIons

As with all new products, you must exercise caution until you become 

proficient in its use. While we have implemented safeguards against 
possible misuse, there are several important points to remember:

1.  A slow-speed handpiece is required for rotary file use.
2.  Operate the handpiece at 300 RPM (revolutions per minute), or  

500 RPM as directed for specific files.

3.  Straight-line access is a prerequisite for proper endodontic treatment. 

ProFile‚ GT files are no exception.

4.  Always utilize minimal apical pressure. Don’t force the files down  

the canal.

5.  Clean file flutes frequently during instrumentation (after the file is 

removed from the canal).

2

6.  Frequently irrigate and lubricate the canal throughout the procedure.
7.  Exercise caution in the apical area and around significant 

curvatures.

8.  Files are single use devices. They can become less efficient after 

multiple uses causing undue stress to the file. This can lead to file 
separation.

9.  When instrumenting the canal, select appropriately sized 

instruments as choosing an overly large file can lead to dangerous 
over-enlargement of the coronal portion of narrow root forms. 
Additionally, too large a file taken to length increases the risk of 
separation.

10.  To reduce the risk of electric shock, do not remove oven cover; there 

are no user-serviceable parts inside.

11.  This oven is intended for use in heating gutta-percha coated 

obturators only. Do not insert anything other than an obturator into 
the oven.

aDverse reactIons

Patients with latex sensitivity may experience allergic reactions to gutta-

percha, which contains dry natural rubber. 

 

steP-by-steP InstrUctIons

Turning on Safe and Efficient Power

ProFile

®

 GT rotary instruments are used in a controlled, slow-speed rotary 

handpiece, thus providing the reaming action that produces a quicker, 
centered, tapered preparation, while maintaining good tactile sense. A 
torque control motor, such as the one included in your system, should be 
used.

Procedures for Using the Aseptico ITR Motor

These are the torque settings that work best with each file series. They are 

given without guarantee of risk or fracture. To avoid separation, remember 
torque control motors are not a substitute for proper technique.

1

  File series 

torque setting

  GT accessory files 

Max

  Coronal 1/3 

3

  Midroot 

2

  Apical third 

1