7 Electromagnetic compatibility
7.2 Interference immunity
66 90 643 D3762
D3762.201.01.03.02 04.2021
31
Ports of signal input/signal output units (SIP/SOP)
Phenomenon
Basic EMC standard
Interference immunity test level for profes-
sional healthcare facilities
Electrostatic discharge
e
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
e
Connectors shall be tested per 8.3.2 and Table 4 of IEC
61000-4-2:2008. For insulated connector shells, perform air discharge
testing to the connector shell and the pins using the rounded tip finger
of the ESD generator, with the exception that the only connector pins
that are tested are those that can be contacted or touched, under
conditions of intended use, by the standard test finger shown in Figure 6
of the general standard, applied in a bent or straight position.
Test specifications for enclosure port immunity to RF wireless
communications equipment
Test fre-
quency (MHz)
Frequency band
a
(MHz)
Radio service
a
Modulation
b
Power
max. (W)
Distance
(m)
Immunity test
level (V/m)
385
380 - 390
TETRA 400
Pulse modulation
b
18 Hz
1.8
0.3
27
450
430 - 470
GMRS 460,
FRS 460
FM
c
± 5 kHz stroke
1 kHz sinus
2
0.3
28
710
745
780
704 - 787
LTE band 13,
17
Pulse modulation
b
217 Hz
0.2
0.3
9
810
870
930
800 - 960
GSM
800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE band 5
Pulse modulation
b
18 Hz
2
0.3
28
1720
1845
1970
1700 - 1990
GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE band 1,
3, 4, 25;
UMTS
Pulse modulation
b
217 Hz
2
0.3
28
2450
2400 - 2570
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE band 7
Pulse modulation
b
217 Hz
2
0.3
28
5240
5500
5785
5100 - 5800
WLAN
802.11 a/n
Pulse modulation
b
217 Hz
0.2
0.3
9
NOTE If necessary, the distance between the transmitting antenna and
the UNIT can be reduced to 1 m in order to achieve the immunity test
level. The 1 m test distance conforms to IEC 61000-4-3.