14

6. Cavitron DualSelect

 

Dispensing system

7. Steri- MATE

®

 Grip

8. Steri- MATE

® 

Light Lighted Sleeve

8 6 2 User Replaceable Part Kits

  

  1.  Cavitron Insert Replacement O-ring Kits, 12/packs 

Part Number 62351 (black) for plastic and soft grips

 Part Number 62605 (green) for metal grips

  2.  Handpiece Cable O-Ring, Part Number 79357

  3.   Lavage (Water) Filter, 10/Pack, Part Number 90158

For detailed information, contact your local DENTSPLY 

Professional Representative or authorized DENTSPLY 

Professional Distributor.

SECTION 9: System Setup, 

Operation and Techniques 

for Use

9 1 handpiece Setup

•  This handpiece is sterilizable. Refer to Infection Control 

Information Booklet for sterilization instructions prior to 

using handpiece.

•  Connect the Handpiece to the Cable Assembly by 

aligning the electrical connections.  If Cable Assembly 

does not seat into the handpiece, gently rotate 

the handpiece until contacts align, then fully insert 

handpiece.

•  Hold empty handpiece in a semi-upright position over a  

sink or drain.  Activate the Tap-On

TM

 Foot Pedal until 

water exits to release any air bubbles that might 

be trapped inside the handpiece. NOTE: Tap-On

TM

 

Technology only operates when an insert is in the 

handpiece.

•  Lubricate the O-ring on the insert with water before  

 placing it into the handpiece.  Fully seat insert with a  

gentle push-twist motion.  DO NOT FORCE. 

•  Turn the Lavage Control to select flow rate during  

system operation. Flow rate is placed on a scale from 1 

to 6. Turn control clockwise toward 6 to increase flow 

at insert tip. Turn control counter-clockwise toward 1 

to decrease flow. The flow rate through the handpiece 

also determines the temperature of the lavage. Lower 

water flow rates produce warmer lavage. Higher flow 

rates produce cooler lavage. If the handpiece becomes 

8 5 Tap-On

 Technology wireless Foot 

Pedal Operation

Using Foot Pedal in Tap-On

T M 

 Mode

For scaling procedures,Tap-On

TM

 Technology eliminates the 

need to hold the pedal down. Tapping the foot pedal once 

activates ultrasonic power or rinse mode for approximately 

4 minutes. Tapping the foot pedal while in Tap-On

TM   

mode 

disables the ultrasonic power and water flow. Boost is still 

available while scaling in Tap-On

TM 

mode. To use boost, simply 

depress the foot pedal to the second position (all the way to 

the floor) to activate and hold as long as boost is desired. 

Release foot pedal to return to Tap-On 

T M 

 mode.

 

TIPS: 

Tap-On

TM

 Technology will not run water unless an 

insert is in the handpiece
A sensor in the handpiece holder will prevent 

Tap-On

TM

 Technology from operating when the 

handpiece is in the holder.
If the foot pedal is not tapped quickly, it will 

function in a conventional manner.

How to Disable and Enable Tap-On

TM

 Technology 

Tap-On

TM

 Technology feature can be disabled by 

simultaneously holding the Purge and Turbo buttons for a 

period of approximately 5 seconds. The two buttons will blink 

approximately 6 times. When the buttons are released, they 

will blink an additional 6 times to confirm Tap-On

TM 

has been 

disabled.
Tap-On

TM

 Technology feature can be enabled by 

simultaneously holding the Purge and Turbo buttons for a 

period of approximately 5 seconds. The two buttons will blink 

approximately 6 times to confirm Tap-On

TM 

has been enabled.  

Using Foot Pedal without Tap-On

T M 

 Mode

For scaling operation, the first position activates both the 

ultrasonic energy and lavage at the insert tip.  The second 

position activates the Boost Mode. The Boost Mode (fully 

depressed Tap-On

TM

 Foot Pedal) increases the ultrasonic 

power level for quick removal of tenacious deposits without 

adjusting the power level knob. To deactivate Boost Mode, 

release Tap-On

TM

 Foot Pedal to first position. 

8 6 Accessories and User 

      Replaceable Parts

 

8 6 1 Accessories

1. AC Power Cord 

2. Tap-On

TM

 Technology Wireless Foot Pedal 

3. Auxiliary Tap-On

TM

 Foot Pedal Power Cable

4. Cavitron Steri- MATE

®

 Sterilizable Handpiece

5. Cavitron 30K Ultrasonic Inserts

NON-DEPRESSED

DEPRESSED 

1st POSITION

DEPRESSED 

2nd POSITION

Summary of Contents for Cavitron Plus

Page 1: ...n TM Tap On TM tecnolog a Tap On TM tecnologia Tap On TM Tap On TM Technologie Technologie Tap On TM Tap On TM tecnolog a Tap On TM tecnologia Tap On TM Cavitron Plus Ultrasonic Scaler Please read car...

Page 2: ...p On Technology does not function unless an insert is in the handpiece TIP Handpiece is detachable and autoclavable Make sure the electrical connections are dry before connecting Place ULTRASONIC INSE...

Page 3: ...e foot pedal all the way to the floor enables quick removal of tenacious calculus Tip Quick release of pedal from Boost will not deactivate Tap On Technology Purge Remove insert from Handpiece and pre...

Page 4: ...E Illuminates when the system is not functioning properly This display has three distinct modes Slow blink 1 blink per second means the system is not operating within factory specifications Fast blink...

Page 5: ...gnostic Display Indicators and Controls 12 8 3 Handpiece Cable 13 8 4 Cavitron 30K Ultrasonic Inserts 13 8 5 Tap OnTM Wireless Foot Pedal Operation 14 8 6 Accessories and User Replaceable Parts 14 8 6...

Page 6: ...e a Tap OnTM Wireless Foot Pedal with Tap On Technology Turbo Mode illuminated diagnostic display rinse setting automated purge function Steri Mate detachable sterilizable handpiece and 330 swivel han...

Page 7: ...0601 Medical Device Standards Insufficient water flow could result in elevated water and tip temperature When operated at the input water temperature specified in the Water Line Requirements Section 7...

Page 8: ...itron Plus Ultrasonic Scaler will create aerosols Following the procedural guidelines in Section 9 of this manual can effectively control and minimize aerosol dispersion 6 2 Water Supply Recommendatio...

Page 9: ...it in direct sunlight may discolor plastic housing The system has been equipped with a Tap OnTM Wireless Foot Pedal which was factory synchronized to operate with the system s base unit If your office...

Page 10: ...the following steps to synchronize the Tap OnTM Foot Pedal with the base unit 1 Turn the MainPower switch located at the center front underside of the system to the OFF O position 2 Install a new set...

Page 11: ...n subgingivally scaling Rinse Turn the ultrasonic power level control knob fully counterclockwise until a click is heard Rinse mode is for use during an ultrasonic scaling procedure when lavage is des...

Page 12: ...cator Illuminates when the Boost Mode has been activated with the Tap OnTM Foot Pedal Turbo Mode When pressed extra power is delivered to the system up to 25 This will increase the stroke of the tip T...

Page 13: ...comfortable grasp of the handpiece The grip is sterilizable and is available in several different colors as an accessory for your Steri Mate Handpiece See installation instructions provided with the g...

Page 14: ...chnology eliminates the need to hold the pedal down Tapping the foot pedal once activates ultrasonic power or rinse mode for approximately 4 minutes Tapping the foot pedal while in Tap On TM mode disa...

Page 15: ...avitron Ultrasonic Inserts are intentionally rounded so there is minimal danger of tissue laceration with proper ultrasonic scaling technique Whenever the insert tip is placed in the patient s mouth t...

Page 16: ...ect the handpiece cable for any breaks or tears 7 If using a closed water supply or DualSelect Dispensing System check for adequate fluid volumes for the next patient 8 When ready to use place a steri...

Page 17: ...Tap OnTM Foot Pedal to the first position The system should dispense water If none of the Tap OnTM Foot Pedals operate the system continue to the next step 2 Resynchronize one Tap OnTM Foot Pedal to t...

Page 18: ...ry Certified Service at 1 800 989 8826 Monday through Friday 8 00 A M to 5 00 P M Eastern Time For areas outside the U S contact your local DENTSPLY Professional representative Section 12 Warranty Per...

Page 19: ...tion of ingress of water Consult Instructions for Use O I AC Power Switch 0 Off On Footswitch MEDICAL EQUIPMENT WITH RESPECT TO ELECTRIC SHOCK FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60...

Page 20: ...61000 3 2 Class A Not Applicable Voltage fluctuations flicker emissions IEC 61000 3 3 Not Applicable The Ultrasonic Scaler is suitable for use in all establishments including domestic establishments a...

Page 21: ...ite surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the high...

Page 22: ...ectromagnetic propagation is affected by absorption and reflection from structures objects and people Recommended separation distance between Portable and mobile RF communications equipment and the mo...

Page 23: ...blinking stops the system is ready for use If blinking remains continue to the next step B Attach a NEW handpiece and depress Tap OnTM Foot Pedal for 2 seconds If blinking stops the system is ready f...

Page 24: ...riv e d air Installez les PILES dans la p dale sans fil Ins rez l insert ultrasonique dans la pi ce main R glez le niveau de puissance R glez la commande de lavage sur la pi ce main au niveau d sir Co...

Page 25: ...cette fonction mains libres permet d augmenter la puissance du syst me pour retirer rapidement les d p ts tenaces il suffit d appuyer sur la p dale jusqu au bout vers le sol Astuce Le mode Tap OnTM re...

Page 26: ...et indicateur peut s allumer de trois fa ons Clignotement lent 1 clignotement par seconde lorsque le syst me ne fonctionne pas selon les param tres d finis en usine Clignotement rapide 3 clignotements...

Page 27: ...me 34 8 2 Voyants et commandes du tableau de diagnostic 35 8 3 Pi ce main c ble 36 8 4 Inserts ultrasoniques Cavitron 30K 36 8 5 Fonctionnement de la p dale Tap OnTM 37 8 6 Accessoires et pi ces remp...

Page 28: ...Zone bleue laquelle permet une plus grande pr cision dans les param tres d alimentation Les fonctionnalit s volu es qui font du Cavitron Plus un investissement judicieux sont la p dale sans fil Tap O...

Page 29: ...e sur un stimulateur cardiaque n ait t rapport DENTSPLY nous recommandons que les pi ces main et les c bles soient maintenus une distance d au moins 15 23 cm de tout appareil implant et de ses fils pe...

Page 30: ...dicateur d efficacit d insert DENTSPLY Professional est inclus pour la commodit de l utilisateur Si une usure excessive est observ e si l insert a t pli ou d form ou s il pr sente d autres dommages je...

Page 31: ...mati re d lectricit L alimentation lectrique du syst me doit amener un courant secteur monophas sous 100 240 volts de 50 60 Hz pouvant atteindre 1 0 amp re L alimentation lectrique doit arriver par l...

Page 32: ...nTM Retournez la p dale sans fil Tap On et enlevez doucement la vis du couvercle du compartiment des piles l aide d un tournevis pointe cruciforme puis retirez le couvercle S il y a lieu enlevez les p...

Page 33: ...es sont allum s appuyez sur le bouton Purge lui m me situ sur le tableau de diagnostic Les voyants se mettront clignoter dans un ordre s quentiel qui refl te le mode de synchronisation Cette phase dur...

Page 34: ...om tre dans le sens inverse des aiguilles d une montre jusqu ce que vous entendiez un clic Le mode de rin age est utile durant une proc dure de d tartrage ultrasonique lorsque vous d sirez un lavage a...

Page 35: ...S allume lorsque le mode Booster est activ l aide de la p dale Tap OnTM Mode Turbo Appuyez sur le bouton pour envoyer jusqu 25 de puissance en plus au syst me Cela allongera la course de l embout Le m...

Page 36: ...lisable elle est disponible en plusieurs couleurs assorties votre pi ce main Steri Mate Veuillez consulter les instructions d installation fournies avec la poign e Les divers types d inserts ultrasoni...

Page 37: ...e enfonc e Le fait de tapoter sur la p dale une fois active le mode de d tartrage ultrasons ou de rin age pendant environ 4 minutes Lorsque la p dale est en mode Tap On tapoter la p dale de nouveau po...

Page 38: ...cadran et la surface en cours de traitement vacuez l eau d irrigation au moyen soit d un jecteur de salive soit d un HVE High Volume Evacuator 9 5 Ex cution des proc dures de d tartrage par ultrasons...

Page 39: ...nouveau ou appuyez sur la p dale Tap On 6 Lorsque vous avez termin le cycle de purge placez un insert Cavitron 30 kHz st rilis dans la pi ce main puis r glez la commande de puissance ultrasonique et l...

Page 40: ...du tuyau rel chez la pression d eau en appuyant sur la pointe du raccord et laissez l eau se vidanger dans un r cipient appropri 3 Saisissez le raccord sur l un des deux c t s du disque filtrant et to...

Page 41: ...n enlevez le Assurez vous que la pi ce main est correctement enfonc e et tenez la p dale appuy e pendant 2 secondes Si le clignotement s arr te le syst me est pr t l emploi Si le clignotement persiste...

Page 42: ...tinue 100 240 V c a Courant 1 0 amp re maximum Phase Simple Fr quence 50 60 Hertz Pression d eau 20 40 psig 138 275 kPa Temp rature de l eau 25 C 77 F Pression d air 65 100 psig 448 600 kPa D bit d ea...

Page 43: ...Consulter le mode d emploi O I Interrupteur de courant alternatif 0 teint Allum Interrupteur au pied quipement m dical concernant les risques de choc lectrique d in cendie et d anomalie m canique uniq...

Page 44: ...de tension scintillements IEC 61000 3 3 Non applicable Le d tartreur ultra sons peut tre utilis dans tous les types de b timents notamment les constructions domestiques et celles qui sont reli es dire...

Page 45: ...ences b Des interf rences sont possibles aux alentours de l appareil elles sont identifi es avec ce symbole REMARQUE 1 A 80 MHz et 800 MHz la gamme de fr quence sup rieure s applique REMARQUE 2 Il se...

Page 46: ...metteurs dont la puissance maximum en sortie n est pas mentionn e ci haut la distance de s paration d recommand e en m tres m peut tre estim e l aide de l quation applicable la fr quence de l metteur...

Page 47: ...tape suivante B Branchez une pi ce main NEUVE et appuyez sur la p dale Tap On pendant 2 secondes Si le clignotement s arr te le syst me est pr t l emploi Jetez l ancienne pi ce main ou retournez la au...

Page 48: ...800 989 8826 or 717 767 8502 De utschland France Australia DENTSPLY DeTrey GmbH DENTSPLY DeTrey DENTSPLY Australia Pty Ltd De Trey Strasse 1 17 Michael FARADAY 11 21 Gilby Road 78467 Konstanz 78380 M...

Reviews: