Symptom/Cause
Corrective Measures
Ultrasonic power level cannot
be changed by sliding finger
along the touch screen:
1. If the screen is dimmed, it
may be locked.
1. Depress padlock icon at the lower left of the main screen to unlock.
Ultrasonic Preset Power
modes cannot be accessed:
1. Preset Power icon not
pressed properly.
1. Press center of Preset Power icon for positive activation.
Ultrasonic Preset Power
Modes cannot be changed:
1. Power preset icons were not
programmed properly.
1. Set desired power level by sliding the power level icon along the scale or tapping
the scale where desired.
2. Press and hold the selected memory icon.
3. Wait for the audible signal. The power level should be indicated in the memory
location icon.
Purge mode cannot be
activated:
1. Purge Icon not depressed
properly.
2. Foot Control is depressed.
3. Handpiece Lavage Control
set too low.
1. Be sure Purge Icon is depressed until activation. Screen will indicate the remaining
Purge time and return to Main screen when completed.
2. Foot Control must be released to activate Purge function.
3. Turn Lavage Control to maximum setting for better purge water flow. Inserts
should also be removed.
Rinse mode cannot be
activated:
1. Foot Control not depressed.
1. Foot Control must be depressed to begin the Rinse mode. The screen will indicate
when Rinse mode is in operation.