6.6.4. Changing Torque and Speed
NOTE
Speed and torque cannot be changed for reciprocating systems.
When the desired continuous rotary file is selected (indicated by an arrow), press the
"
key to
move to the torque field, then indicated by an arrow. Use the
+
and
–
keys to select the desired
torque setting. When the torque value is changed, it will flash. The available torque range is
0.2 to 4.0 Ncm in steps of 0.1 Ncm.
After selecting the desired torque, save the setting by pressing the
OK
key once.
The torque field will then stop flashing and will be marked with the
‘
symbol. If the
OK
key is
not pressed to save the setting, the setting cannot be used and will be lost once a different file
setting is selected.
Press the
"
key to move to the speed field, then indicated by an arrow. Use the
+
and
–
keys
to select the desired speed setting. When the speed value is changed, it will flash.
The available speed range is 250 to 1000 rpm in steps of 10 rpm (with the provided
WaveOne
!
(MAILLEFER 6:1) contraangle only).
The preset torque and speed values of all continuous rotary file systems can be altered
individually.
WARNING
Before using the micromotor, verify the correctness of the changed parameters.
6.7.
Factory Default Parameters
To return to the original default parameters, follow the general reset instructions:
!
Make sure that the battery charger is disconnected.
!
Switch off the device.
!
Keep holding down the
OK
key and the
POWER
key simultaneously. The device will switch
on and the display will read: