![DENTIS LUVIS L200 User Manual Download Page 15](http://html1.mh-extra.com/html/dentis/luvis-l200/luvis-l200_user-manual_2483478015.webp)
Professional LED Light System for All of Surgery Application
15
Ÿ
Recommended separation distances between portable and mobile RF
communications equipment and the L200
The L200 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the L200 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF communications equipment (transmitters) and the L200 as recommended below, according
to the maximum output power of the communications equipment.
Rated
maximum
output
power of
transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
=
3,5
1
√ = 1.17√
80 MHz to 800 MHz
=
3,5
1
√ = 1.17√
800 MHz to 2,5 GHz
=
7
1
√ = 2.34√
0,01
0,12
0,12
0,23
0,1
0,37
0,37
0,74
1
1,17
1,17
2,33
10
3,69
3,69
7,38
100
11,67
11,67
23,33
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.