of the apical constriction. (Generally we suggest to use 0.5 for 
measurement the length of root canal). 
Note: Please do not make the measurement when in charge.   
6.

 

Deciding the working length of root canal 

      Measure the distance from the bottom of rubber positioning ring to 
the tip of the file until the figure indicated 0.5. So the length of root 
canal need to minus 0.5~1.0mm is the most suitable working length of 
root canal. 
   The working length of root canal varies from different shapes of 
teeth and root canal. 
7.

 

After operation, please pull out the probe wire and switch off the 

instrument. If the dentists forgot to switch off the instrument, the 
instrument will automatically shut down. 
 

SECTION 6: SAFETY PRECAUTIONS 

     

 

CAUTIONS: 

      a. Before operation, you have to read user manual carefully. 
   b. Like all of the other electric facilities, this machine has an 
electromagnetism disturbance. When there is a patient who is now 
using the cardiac pacemaker, or there is an electronic operation, 
please don't put the machine around. The cardiac pacemaker 
sufferer, viz. the serious cardiac pulse abnormality sufferer, is forbidden 
to use this machine. 
      c. Please put in the battery before use. Make sure that the power 
of the battery is in sufficient supply to guarantee the correct 

                               

         

VER DY20130812RPIF-EN   

16

 

measurement result.   
      When change the battery, do not mix the old battery with the new 
one and mix the alkali battery with the manganic one. 
      Please take off the battery in the event of longtime nonuse or long 
–distance transit.   
      d. Please use the file with the resin handle rather than metal one. 
Even when using the file with the resin handle, please notice that the 
fingers should be avoided touching the mental part of the file.   
   e. Please clip the upper portion of the file rather than the down 
portion with the holder, other wise, the metal part of the holder and 
the resin part of the file would be damaged. The damaged holder will 
affect the measure result.   

 

  f. When the file accidentally touches the inner part of the 

root-canal, the reading of scale will get a bit abnormality, then will get 
right automatically a few seconds later. 

 

      g. The device is not suitable for use in the presence of flammable 
anesthetic mixtures with air or with oxygen or nitrous oxide. 
      h. The enclosure of the main body of device is not designed to give 
any protection against ingress of water. Please keep the device away 
from any harmful ingress of water. 
 

For ACCURATE MARESUREMENT: 

 Make sure that stainless hook entirely contact patient’s mouth 

mucosa 

 Check all connections 

 Make sure that when the device is switched on, the device can 

Summary of Contents for iFinder

Page 1: ...RMATION Apex Locator iFinder is manufactured by DENJOY DENTAL CO LTD Address 4th Floor A3 Building Lugu International Industry Park No 229 Tongzipo Road West Changsha City P R China 410205 Website www...

Page 2: ...angle Up to date multi frequency operating system Rechargeable do not need prepare extra battery III SYMBOL DESCRIPTIONS The following symbols may appear in this manual on the label or on it s accesso...

Page 3: ...5V 1A Battery 3 7V 2200mAh 3 Display mode 4 3 inch TFT 480 272 color screen 4 Dimensions 132 120 31mm 5 Weight about 500g 6 Classification Internally powered equipment 7 Degree of protection against...

Page 4: ...mation OK key 10 Length between top of file and the apex of root canal 11 Scale displaying the magnified apex area 12 Apical Area Adjustment cursor a Save key VER DY20130812RPIF EN 8 On the Main Inter...

Page 5: ...uch sound adjuster user can select required volume level after adjustment touch enter key if as shown in Figure 4 it means set up successfully Please note no operation for 10 seconds it will automatic...

Page 6: ...n Figure 5 it means self checking successfully Otherwise if as shown in Figure 6 it means self checking unsuccessfully VER DY20130812RPIF EN 12 Figure 5 Figure 6 b Calibration key for accessories prob...

Page 7: ...wire and insert the stainless electrode hook into the socket 3 Extended pressing power on off button Note Extended pressing power on off button for 2 seconds to turn on off the device 4 Clip the meta...

Page 8: ...at the power of the battery is in sufficient supply to guarantee the correct VER DY20130812RPIF EN 16 measurement result When change the battery do not mix the old battery with the new one and mix the...

Page 9: ...ON 7 TOUBLESHOOTING GUIDE Question After switch on the machine the LCD screen has no reaction Answer a Check that the machine if it need to be changed VER DY20130812RPIF EN 18 Question No alert Answer...

Page 10: ...INSTRUCTION When the surface of main body is polluted please rub the surface with dry soft cloth ONLY REMARKS Any liquid lotion like ethanol banana oil and light oil are not allowed PROBE WIRE CLEANIN...

Page 11: ...VER DY20130812RPIF EN 21 WARRANTY REGISTRATION FORM Item Name Model Name Serial No Date of Purchase Name Address Phone Email Name of Distributor Authorized Distributors Stamp and Signature...

Reviews: