Deltec CADD-Prizm PCSII 6101 Technical Manual Download Page 24

21

Mechanical Inspection

• Press each key on the keyboard. Each key

should have a distinctive dome feeling. The
keys should not feel flat.

• Attach the battery door. The battery door

should fit snugly in place when it is closed on
the pump.

• Attach either a 50- or 100-ml Medication

Cassette reservoir or a CADD

®

 administra-

tion set to the pump. Check for smooth
operation and a definite “feel” when the latch
pulls the reservoir or administration set firmly
against the bottom of the pump. The mark on
the latch should be aligned with the solid dot.

• Lock the device by inserting a key into the

lock and turn counterclockwise until the
mark lines up with the solid dot.

• Gently twist and pull on the cassette to make

sure it is firmly attached.

NOTE:

In the PCA delivery mode, the cassette must
be locked in order to start the pump.

Summary of Contents for CADD-Prizm PCSII 6101

Page 1: ...32 T E C H N I C A L M A N U A L CADD P r i z m P C S I I AMBULATO RY I N F U S I O N P U M P M O D E L 6 1 0 1 Deltec...

Page 2: ...operation of CADD pumps please call this num ber 800 426 2448 Our staff is available to help you 24 hours a day with the programming and operation of CADD pump infusion systems The issue date of this...

Page 3: ...chdog Circuit 12 Power Circuitry 13 Voltage Reference Circuit 13 Pumping Mechanism 13 Pumping Characteristics 14 Air Detector 15 Upstream Occlusion Sensor 15 6 Safety Features and Fault Detection 16 H...

Page 4: ...ED WARRANTY OF MERCHANTABILITY OR FITNESS FOR USE Deltec further disclaims responsibility for the suitability of the system for a particular medical treatment or for any medical compli cations resulti...

Page 5: ...e infusion Epidural administration is limited to short term infusion of anesthetics and either long or short term infusion of analgesics Subarachnoid administration is limited to short term infusion o...

Page 6: ...loading dose Demand Doses Continuous Rate Time Dosage PCA Delivery Profile The PCA patient controlled analgesia delivery mode is used for therapies that require a continuous rate of infusion patient c...

Page 7: ...92 0 4 00 1600 85 4 25 841 5 4 25 1700 90 4 50 891 0 4 50 1800 95 4 75 940 5 4 75 1900 100 5 00 990 0 5 00 2000 Table 3 Demand dose clinician bolus scroll ranges micrograms Demand Dose Clinician Bolus...

Page 8: ...lf life temperature conditions delivery rate and frequency of screen display backlighting and printing It is recommended that a new 9 volt battery be kept available for replacement if necessary An int...

Page 9: ...micro grams mcg Default milligrams Time 00 00 to 23 59 Air Detector Turned On or Turned Off Default On Biomed Toolbox Specifications Custom Concentrations All individual mg or mcg concentration setti...

Page 10: ...e Remote Dose cord may be fastened to the patient s clothing or bedsheet with the attached clip NOTE To detach the Remote Dose cord from the pump grasp the Remote Dose cord connector and pull back usi...

Page 11: ...be shorter if the battery is stored above room temperature An alkaline battery stored at 43 C 110 F will be down to approximately 80 of its capacity within one year Recommended storage conditions are...

Page 12: ...60 hrs 3000 ml 100 ml hr 30 hrs 3000 ml 200 ml hr 32 hrs 6400 ml 350 ml hr 17 hrs 5950 ml Table 4 9 volt Alkaline type batteries used with the CADD Prizm pump Table 5 9 volt Lithium type batteries us...

Page 13: ...oard keys are sensed by the pump s microprocessor The custom Liquid Crystal Display LCD also located on the front housing shows the pump status and programmed settings The dot matrix display consists...

Page 14: ...and record keeping parameters are stored in a battery backed Random Access Memory RAM Battery backup is provided by two printed circuit board mounted lithium batter ies These batteries are designed t...

Page 15: ...a rate of 0 8 and 2 Hz Low battery voltage detection and watch dog timer circuitry also have the ability to enable the audible alarm via the Gate Array Watchdog Timer Circuit Watchdog timer circuitry...

Page 16: ...sely regulated and filtered 5 volt power which is supplied from the micropower voltage regula tor This regulator will supply 5 volt power until its input voltage is approximately 5 3 volts After that...

Page 17: ...mum pressure which can be developed is 186 kPa 1 86 bar 27 psi To deliver the amount of drug specified by the parameter settings the pump s microprocessor causes the pump mechanism to deliver 0 05 ml...

Page 18: ...mpartment with a detachable door It is sealed against the pump housing to ensure the overall assembly is water resistant The air detector is mounted to the pump housing with two screws and electricall...

Page 19: ...st strobe the watchdog circuit at least once every second in order to prevent the watchdog from performing its reset function The reset output from the watchdog circuit is a pulse output This acts to...

Page 20: ...ed states thus allowing detection of sensor fault conditions Care should be taken not to damage these sensor pins Latch Lock Sensor Circuit Latch and Lock sensors allow the microproces sor to detect t...

Page 21: ...tware will turn on a continuous two tone audible alarm and stop all drug delivery The DOSE key has two independent signal lines to prevent single fault failures Data Handling Software Safety Features...

Page 22: ...Order of Error Code Events 1 There is a continuous two tone audible alarm a continuous amber indicator light and the display will read NOTE XXXXX is a 5 digit code 2 To silence the error code alarm r...

Page 23: ...pushed in straightening may be possible with a small screwdriver or other suitable tool Care must be taken so as not to damage the pump housing or to incur further damage to the contacts Inspection R...

Page 24: ...nistra tion set to the pump Check for smooth operation and a definite feel when the latch pulls the reservoir or administration set firmly against the bottom of the pump The mark on the latch should b...

Page 25: ...or to the pump reviewing the current program settings the pump has experienced an electrical or mechanical fault and should be returned for service If no error message is immediately shown the pump ha...

Page 26: ...s the ENTER key Select the 4 hour limit to No Limit then press the ENTER key Select the Reservoir Volume of 2 0 ml then press the ENTER key Motor and Gear Train Check Attach either a 50 or 100 ml Medi...

Page 27: ...RUNNING appears on the display press the Remote Dose Cord button and note the time The pump should beep twice and begin to deliver Count the number of pump activations The pump should make ten double...

Page 28: ...ached filled clamped Medication Cassette reservoir The pump is started and a Demand Dose is given until the high pressure alarm sounds Equipment needed 50 or 100 ml Medication Cassette reservoir conta...

Page 29: ...cifications section for the pump That is under the test conditions described below the accuracy of the pump and reservoir will be nominal with a 90 confidence level The nominal test conditions are as...

Page 30: ...voir extension set and weight of the water 5 Attach the Medication Cassette reservoir to the pump Program the reservoir volume to 20 ml Now press the ENTER key This value is the intended delivery volu...

Page 31: ...anti siphon valve A fluid collection device such as a burette or a class A 25 ml capacity graduated cylinder 40 ml of room temperature water Procedure 1 Fill the 50 or 60 ml syringe with 40 ml of wate...

Page 32: ...5 x 100 2 5 Table 15 Volumetric percentage calculation 8 Divide the inaccuracy volume by the intended delivery volume and multiply by 100 This is the accuracy error percentage See Table 15 9 If the ac...

Page 33: ...mpartment III Mechanical Inspection Keyboard Cassette Latch Battery Door Cassette Lock IV Functional Inspection Power up RES VOL Alarm Dose Given Attempted Latch Lock Stop Start MG Given Cassette Sens...

Page 34: ...O is a LEMO USA INC trademark SaniCloth is a PDI trademark The products described are covered by one or more of the following U S Patent Nos 4 559 038 4 565 542 4 650 469 5 181 910 5 338 157 5 364 242...