21
Mechanical Inspection
• Press each key on the keyboard. Each key
should have a distinctive dome feeling. The
keys should not feel flat.
• Attach the battery door. The battery door
should fit snugly in place when it is closed on
the pump.
• Attach either a 50- or 100-ml Medication
Cassette reservoir or a CADD
®
administra-
tion set to the pump. Check for smooth
operation and a definite “feel” when the latch
pulls the reservoir or administration set firmly
against the bottom of the pump. The mark on
the latch should be aligned with the solid dot.
• Lock the device by inserting a key into the
lock and turn counterclockwise until the
mark lines up with the solid dot.
• Gently twist and pull on the cassette to make
sure it is firmly attached.
NOTE:
In the PCA delivery mode, the cassette must
be locked in order to start the pump.