the logger to the outside of the unit. Allow at least 1 hour for the sensor to reach 
equilibrium before activating the logger. When the sensor doesn’t have enough 
time to equilibrate to the internal temperature, it will trigger a false alarm if it is still 
above the high alarm set point. 

Starting the Logger

Press and hold the 

START

 button until a temperature reading appears on the 

LCD without flashing. If the glycol bottle is inside the storage unit and was 
allowed to acclimate for at least 1 hour, it will be simulating the temperature of 
the vaccines vials stored around it. If 

LLL.L

 appears on the 

LCD, the sensor is not connected. Check the stereo jack to 
make sure it is completely inserted.

Daily Statistics on LCD

Temperature statistics from the last 24 hours of recorded data are available for 
viewing on-demand at any time without stopping the logger. 
A single press of the green 

START

 button shows a statistic on the LCD, each 

subsequent button-press advances to the next statistic.  The LCD automatically 
reverts to current reading about five seconds after button release and showing a 
statistic. The LCD scrolls through values in this order: 

• 

Maximum temperature

• 

Minimum temperature

• 

Average temperature

If alarms limits are programmed:

• 

High alarm set point

• 

Accumulated time above high alarm, if no excursions, 00:00 is displayed

• 

Low alarm set point

• 

Accumulated time below low alarm, if no excursions, 00:00 is displayed

Inserting Waypoints 

If a temperature reading needs to be confirmed at a check point during 
transportation or storage, a waypoint can be inserted as a marker to flag date/
time of this event. The waypoint option must first be enabled during logger 
configuration.  
Press and hold the 

START

 button until 

FLA9 

appears on the 

LCD and release when it stops flashing. The waypoint has 
been inserted and the LCD will revert to current temperature 
reading. When reports are created at the end of the logging 
period, waypoints appear as a vertical line on the graph and have an asterisk in 
the data table.   

Stopping the Logger

I

f monitoring needs to continue after the logger is downloaded, do not remove 

the sensor from the storage unit.

. Press and hold the red 

STOP

 button until 

Summary of Contents for FlashLink 40527

Page 1: ...Quick Start Guide Models 40527 40527 01 Certified Vaccine PDF Data Logger ...

Page 2: ...the license document and note the license key number Double click on FlashLink exe and follow instructions in the dialog boxes that appear on the screen When you get to the Enter License dialog box key in the number provided in the license and select Activate The FlashLink shortcut icon will appear on the desktop and system tray Double click the icon to open the program FlashLink Program Manager r...

Page 3: ... in the table of contents or by using the keyword search bar Plug the FlashLink Data Logger into the USB port of the computer select Setup to open the Logger Configuration Wizard The current configuration is displayed Click on Next to access Logging Options and select desired configuration parameters New Configuration window will confirm settings select Next and OK to upload the configuration to t...

Page 4: ...Glycol Bottle This digital data logger DDL features a stainless steel sensor enclosed in a bottle of propylene glycol The cable has a stereo jack for quick disconnect to easily detach the logger and download data The glycol buffered sensor emulates products in the storage unit and keeps the sensor from setting off false alarms due to rapid temperature fluctuations caused by frequent door opening d...

Page 5: ...statistic The LCD automatically reverts to current reading about five seconds after button release and showing a statistic The LCD scrolls through values in this order Maximum temperature Minimum temperature Average temperature If alarms limits are programmed High alarm set point Accumulated time above high alarm if no excursions 00 00 is displayed Low alarm set point Accumulated time below low al...

Page 6: ... reports using its onboard software and saves these files to an internal memory This allows personnel to download and save reports on any PC and does not require FlashLink Program Manager software Option 2 Download data to a PC which is running FlashLink Program Manager software When a data logger is plugged into the USB port the software automatically detects the logger generates and saves Adobe ...

Page 7: ...stops flashing on the LCD Release the buttons and the LCD will flash all icons while it resets When you see a blank screen with first plug the logger back into the glycol bottle sensor connector and then press and hold the START button until the current temperature reading appears on the LCD The logger will now resume recording Remaining Battery Voltage The LCD includes a battery icon to indicate ...

Page 8: ...741 2837 2664 8923 2778 www deltatrakchina com cn DTI China Limited NT Hong Kong 852 3568 5538 www dtiap com DTI Asia Pacific Auckland New Zealand 64 9 5757 886 www deltatraksouthpacific com DTI South Pacific Osaka Japan 81 6 6616 5900 www dtijapan co jp DTI Japan Limited Antwerp Belgium 32 0 3 455 61 25 www deltatrakeurope be DTI Europe bvba DTI South America SA Santiago Chile 562 2758 2866 569 7...

Reviews: