HD2016
- 16 -
V1.4
Command
Reply
Description
S0
String with 16 fields
Single print of all the measures:
campo 1
⇒
& followed by a control code
campo 2
⇒
Filling percentage of the rain gauge
reservoir
campo 3
⇒
Weight, in mg, of the current reservoir
content (Tmg)
(*)
campo 4
⇒
Rainfall total weight, in mg, considering
the evaporation (Wmg)
(*)
campo 5
⇒
Rainfall total weight (Pmg) in mg
(*)
campo 6
⇒
Partial rainfall, in µm, from the last reset
command
campo 7
⇒
Average rainfall rate, in µm/min, calcu-
lated in the last minute
campo 8
⇒
Average rainfall rate, in µm/h, calculated
in the last hour
campo 9
⇒
Reservoir status (
Evap
if in evaporation,
Rain
if in precipitation) followed by * (va-
lid data) or ! (invalid data)
campo 10
⇒
Indicates how long the status reported in
the previous field has been active
campi 11…15
⇒
Confidential information
campo 16
⇒
Internal temperature in °C
(*)
For details on weight measurements, please see
the paragraph “Notes on weight measurements”.
R0
String with 16 fields
Continuous print (once per second) of all the meas-
ures (see the command S0 for the fields)
ST
address
: &
voltage
status_register
Prints the supply voltage in Volt and the content of
the status register in hexadecimal
SS
address
:
status_register
Prints the content of the status register in hexadeci-
mal