HD2010UC/A
- 55 -
V4.1
ACOUSTIC CALIBRATION
Last:2003/01/01 10.00
Calibr.level 94.0dB
Leq
dBC
Turn off the calibra-
tor.
94.0
NEXT
5.
After the acoustic calibration, the
electric calibration
will be automatically started. This
stage of the procedure generates the reference data for the following electrical calibrations.
ACOUSTIC CALIBRATION
Last:2003/01/01 10.00
Calibr.level 94.0dB
Leq
dBC
Confirm calibration?
Δ
-0.1
ACOUSTIC CALIBRATION
Last:2003/01/01 10.00
Calibr.level 94.0dB
Waiting for electric
calibration ...
Leq
dBC
Δ
12.5
YES
NO
6.
At the end, if the values of the electrical calibration are acceptable, press YES to confirm the
new calibration (LEFT arrow of the keypad); press NO if you want to cancel the whole cali-
bration (RIGHT arrow of the keypad).
7.
Finally, the microphone polarization check is carried out. Wait until "Take out the calibra-
tor" appears (take out the preamplifier from the calibrator".
ACOUSTIC CALIBRATION
Last:2003/01/01 10.00
Calibr.level 94.0dB
Take off the
calibrator
ACOUSTIC CALIBRATION
Last:2003/01/01 10.00
Calibr.level 94.0dB
...
Leq
dBC
Checking microphone
polarization
Δ
-0.1
EXIT
Leq
dBC
96.1
8.
Take out the preamplifier from the calibrator and press EXIT.