Delta Elec. Inc.
Qumi Ver.02
Company Confidential Vivitek ______________________ Delta____________
49
2011/07/05
6-2. Calibrate Analog YUV
Equipment: VG828, must be take off H/V sync terminal from machine
.
Step 1
:
In Service Mode select ADC Calibration - YUV. In this menu input pattern with 75
﹪
SMPTE
pattern for calibrate YUV Source and notice that the input source must be 480i format
。
After input ready press “Enter”key to proceed
。
Input Pattern
Step 2 : After complete the OSD will show the picture as follow, if not complete, just close the OSD
。
NOTE:
You can select Calibration Status – VGA or Calibration Status – YUV item to watch the
calibrated value at any time
。