QUICK GUIDE LTL-X MARK II-US VERS 1.02 JUNE 2015 

 

Calibration

 

There are two kinds of calibrations procedures: 

 

Calibration (red field calibration unit)  

 

Verification (black reference calibration unit). 

      Please refer to the manual for an explanation of these two procedures and when to perform them.  

 
During daily use only a quick 

calibration

 has to be done. There is no need with LTL-X Mark II to make a 

Zero 

calibration

 as was the case with the predecessor LTL-X. The apparatus will guide you through the procedure. 

 

Calibration procedure

 

   

Before a calibration make sure that the calibration unit and front window are clean. 

 
      Calibration using the red field calibration unit: 

 

Mount the calibration unit underneath at the front of the instrument with the white ceramic on the 

calibration unit facing the tower of the instrument. It is important that the pins on the side of the unit 
fit into the slots in the LTL-X Mark II structure.  

Press the

 button and then the 

OK

 button. Check and if necessary correct the calibration value 

shown in the display so it matches the value written on label on the calibration unit. Press the 

OK

 

button to start the calibration. 

  

The calibration procedure is now done. Remove the calibration unit and store it properly. Press 

 

to return to 

the measurement display. 

 
 

Miscellaneous

 

The measure display can at all times be shown by pressing 

             

 

Reset Log:   

Press the 

MENU

 button 

 and select 

LOG/CLEAR DATA/ALL

 
Date and Time: 

Press the 

MENU

 button and select 

SETTINGS/DATE & TIME/SET TIME/DATE

. Select wanted 

value and use ▲ and ▼ to set the time and date. Accept the setting by pressing the green 

OK

 button. Alternatively 

SYNC TO GPS

 can be selected to auto set the time. Remember 

also to set correct time zone. 

       
Power Save:   

Press the 

MENU

 button and select 

SETTINGS/POWER

. Select 

AUTO SLEEP

 

or 

AUTO OFF

 

and use ▲or ▼ to edit the auto turn off / sleep time. Accept the setting by pressing the 
green 

OK

 button. 

 

Sound:  

 

Press the 

MENU

 button 

 and select 

SETTINGS/SOUND

SELECT KEY CLICK

 or 

SOUNDS

 to set the individual sound levels for key click and miscellaneous system sounds. 

Use ▲ or ▼buttons to set the levels. Accept the setting by pressing the 

OK

 button. 

Activate the 

BEEP 

sound to inform when the measurement has been taken 

 

    Warnings and Errors:   An error icon 

 or warning icon 

 will be displayed in the top line of the display 

if a problem is detected when a measurement is taken. Press ▲ to get a description of the 
most severe error or warning. Now press the 

OK

 button to display a total list of errors or 

warnings. 

Reviews: