Other bridge sensors 

Sensor connection 

Connect sensor wires to GP1-PBA1 terminals: 

GND, Vref 

Precision bridge excitation or power supply 

CH(-), CH(+) 

Bridge output 

Step2: Measure the zero offset 

If possible, zero the sensor output and connect the sensor to the GP1-PBA1, otherwise 
fit the Calibration Resistors and wire link, as for SWT5 tensiometers. 
Measure the 

zero offset

,

 

as for SWT tensiometers. 

Step 3: Program the GP1 for bridge sensor readings 

Proceed as for SWT tensiometers. 
In 

Voltage Channel Properties, 

enter appropriate 

Warmup

Units, Minimum, 

Maximum 

and 

Resolution

 values for your sensor.

 

To record sensor readings in mV: enter 

a1

 

= 0.02  

[the GP1-PBA1 gain is 50]

 

For readings in ‘engineering units’, calculate 

a1 

from 

sensitivity

 

and 

excitation voltage

 

values provided by the manufacturer of the sensor. 

Sensitivity

 is typically specified as mV per engineering unit, at a given 

excitation 

voltage

 in volts, e.g. 0.1mV / hPa at 10.6V excitation:  

a1

 = 

excitation voltage

 

/ ( 150 x 

sensitivity

 

)  

[3 volts, times a gain of 50]

 

Alternatively, 

sensitivity

 may be specified in mV per engineering unit per volt excitation, 

e.g. 0.01mV per mbar per volt excitation:  

a1

 = 

1

 

/ ( 150 x 

sensitivity

 

In all cases, calculate and enter 

a0

 = 

– a1

 

zero offset

 

Specification 

Typical contribution to reading errors after zero offset correction, assuming nominal excitation and 
gain values

 

Electrical parameters 

 

Precision

 

bridge excitation: 3.0V

Bridge impedance: >300 

Ώ

 

Input range: ±29 mV  
Gain: 50 
Output offset: 1250 mV nominal 

Accuracy 

 

0 to 40°C: 

±0.1 mV ±0.06% of reading

  

-20 to +60°C: 

±0.2 mV ±0.08% of reading 

SWT tensiometer (0 to 40°C): 

±4 hPa ±0.06% of reading 

Other bridge sensors:

 If common mode changes after 

zero offset measurement add ±0.14 mV / V of CM change 

Notices 

All parts of the GP1 Data Logger and GP1-PBA1 Precision Bridge Adapter designs and 
documentation are the exclusive property of DeltaT Devices Limited, and are covered under 
international copyright law. 

 

 

 

 

Delta-T Devices Ltd 
128 Low Road, Burwell 
Cambridge CB5 0EJ 
UK 

Tel

:  

+44 1638 742922 

Fax

+44 1638 743155 

e-mail

[email protected] 

web

www.delta-t.co.uk

Reviews: