 INTRODUCTION 

LPPYRA-Lite  pyranometer  measures  the  irradiance  on  a  flat  surface  (W/m

2

).  The 

measured irradiance  (Global Irradiance) is the sum of direct solar  irradiance and dif-
fuse irradiance.   

LPPYRA-Lite  is  a  Spectrally  Flat  Class  C  (Second  Class)  pyranometer  in  accordance 
with  ISO  9060:2018  and  with  the  criteria  of  the  WMO  “Guide  to  Meteorological  In-
struments and Methods of Observation’”. 

The pyranometer is available in the following versions: 

 

LPPYRA-Lite

 PASSIVE. 

 

The  passive  version  can  be  connected  to  the  instruments 
D09847 and HD31 by using VP472 SICRAM module. 

 

LPPYRA-LiteAC

:   ACTIVE  with  2-wire  (current  loop)  4..20  mA  output  (0…2000 

W/m

2

). 

The  pyranometer  is  supplied  factory  calibrated.  The  calibration  is  carried  out  in 
accordance  with  the  ISO  9847:1992  (type  IIc)  standard:  ”Calibration  of  field 
pyranometers  by  comparison  to  a  reference  pyranometer”.  The  pyranometer  is 
calibrated by comparison with the reference sample calibrated annually at WRC (World 
Radiation Center). 

An optional fixing base integrating a levelling device is available as an accessory, for 
installations where horizontal positioning is required. 

Summary of Contents for LPPYRA-Lite

Page 1: ...Operating manual Pyranometer LPPYRA Lite www deltaohm com English Keep for future reference...

Page 2: ...LE 4 3 INSTALLATION 5 4 ELECTRICAL CONNECTIONS 8 4 1 LPPYRA LITE CONNECTIONS 8 4 2 LPPYRA LITEAC CONNECTIONS 9 5 MEASUREMENT 10 5 1 LPPYRA LITE 10 5 2 LPPYRA LITEAC 10 6 MAINTENANCE 11 7 TECHNICAL SPE...

Page 3: ...PYRA Lite PASSIVE The passive version can be connected to the instruments D09847 and HD31 by using VP472 SICRAM module LPPYRA LiteAC ACTIVE with 2 wire current loop 4 20 mA output 0 2000 W m2 The pyra...

Page 4: ...ence of temperature between hot and cold junction is converted into a Difference of Potential In order to grant the thermopile a proper thermal insulation from the wind and reduce the sensitivity to t...

Page 5: ...he pyranometer itself In compliance with ISO TR9901 standard and WMO recommendations the pyra nometer must be positioned so that its connector is pointed to the North Pole if the instrument is used in...

Page 6: ...32 LPS40 32BL Integrated levelling device Fig 3 2 fixing adapters from 40 to 32 mm holes centre distance HD2003 77 40 LPPYRA Lite LPS40 32 Fig 3 3 fixing on HD2003 77 40 clamping LPS3 LPPYRA Lite LPS...

Page 7: ...LPPYRA Lite 7 V1 2 LPPYRA Lite LPS40 32 LPRING04 Fig 3 3 fixing on LPRING04 adjustable holder...

Page 8: ...to the housing Grounding the housing allows the correct protection func tionality in particular against lightning 4 1 LPPYRA Lite CONNECTIONS The pyranometer LPPYRA Lite is passive and does not requi...

Page 9: ...d an instrument with 4 20 mA input as shown in fig 4 2 The load resistance of the instrument reading the signal must be 500 Connector Function Color 1 Positive Iin Red 2 Negative Iout Blue 3 Not conne...

Page 10: ...of the sensor and applying the following formula Ee DDP S where Ee is the irradiance expressed in W m2 DDP is the difference of potential expressed in V measured by the multimeter S is the sensitivit...

Page 11: ...ntly the more the dome will be kept clean the more measure ments will be accurate You can wash it using water and standard papers for lens If necessary use pure ETHYL alcohol After using alcohol clean...

Page 12: ...al to output signal 4 20 mA Connection 4 pole M12 connector Dimensions Fig 3 1 Weight 150 g approx Operating temperature humidity 40 80 C 0 100 Protection degree IP 67 MTBF 10 years Technical Specific...

Page 13: ...well as the specific measures described in this manual are followed The instrument proper operation and operating safety can be ensured only in the cli matic conditions specified in this manual Do no...

Page 14: ...r for mounting the pyranometer in an inclined position on 40 mm mast with internal thread LPS40 32 adapter is required HD2003 77 40 Clamping for mast 40 mm to install the pyranometer on a transverse m...

Page 15: ......

Page 16: ...rranty included in the manual of the product For any dispute the competent court is the Court of Padua The Italian law and the Convention on Contracts for the International Sales of Goods apply TECHNI...

Reviews: