Delta OHM HD3406.2 User Manual Download Page 8

-      - 

8

tion year is not displayed. The cell constant change, using the K CELL parameter, resets the 
date. 

8)

 

ALPH_T (Temperature coefficient 

α

T

): 

the temperature coefficient 

α

T

 is the percentage 

measurement of the conductivity variation according to temperature and is expressed in 
%/°C (or %/°F). The admitted values vary from 0.00 to 4.00%/°C. Use the arrows (

 and 

) to set the desired coefficient 

α

T

, and confirm with ENTER. 

9)

 

REF_TEMP (reference temperature)

: it indicates the temperature to which the displayed 

conductivity value is standardized and can be equal to 20°C or 25°C. Using the arrows (

 

and 

), set the desired value and confirm with ENTER. 

10)

 

TDS (Conversion factor 

χ

/TDS)

: it represents the ratio between the measured conductivity 

value and the total quantity of dissolved solids in the solution, expressed in mg/l (ppm) or g/l 
(ppt). This conversion factor depends on the nature of the salts present in the solution: in the 
field of water quality treatment and control, where the main component is CaCO

(Calcium 

Carbonate), a value of 0.5 is usually used. For agriculture water, for fertilizers preparation, 
and in hydroponics, a factor of about 0.7 is used. Using the arrows (

 and 

), set the de-

sired value, selecting it in the 0.4…0.8 range, and confirm with ENTER. 

11)

 

RCD MODE (Record mode)

: the instrument captures a conductivity and a temperature 

value every second. If the RCD MODE parameter is set to "

conductivity

" (factory default), 

the maximum (MAX) and minimum (MIN) values displayed using FUNC refer to conduc-
tivity: the indicated temperature is that measured at the maximum and minimum conductiv-
ity and is not the maximum and minimum temperature.  
If the RCD MODE parameter is set to "

tp

" (=temperature), the maximum and minimum 

values displayed using FUNC refer to temperature: the indicated conductivity is that meas-
ured at the maximum and minimum temperature and is not the maximum and minimum 
conductivity. 
Finally, if the RCD MODE parameter is set to "

Indep

" (=independent), the maximum and 

minimum values displayed using FUNC are independent: the indicated conductivity and 
temperature are the maximum and minimum measured values but are not necessarily re-
ferred to the same measurement moment. 

12)

 

Probe type (

the message "PRBE_TYPE" is scrolled in the comment line. The main line in 

the center of the display shows the type of temperature

 

probe connected to the instrument. 

Conductivity/temperature combined probes with Pt100 or Pt1000 sensor, or temperature 
only probes can be connected to the input:  

 

 4 wire PT100 using the TP47 module 

 

2 wire PT1000 using the TP47 module  

Upon being turned on,

 the instrument automatically detects the temperature probes: the 

Probe Type

 menu item is configured by the instrument and cannot be modified by the user. 

 If no temperature probe or combined probe with temperature sensor is connected, the in-
strument displays a dotted line (- - - -). 

13)

 

YEAR

: to set the current year. Use the arrows to modify this parameter and confirm using 

ENTER. 

14)

 

MNTH (month)

: to set the current month. Use the arrows to modify this parameter and con-

firm using ENTER. 

15)

 

DAY

: to set the current day. Use the arrows to modify this parameter and confirm using 

ENTER. 

Summary of Contents for HD3406.2

Page 1: ...mation written in this manual and the instru ment that you have purchased We cannot entirely exclude errors in the manual for which we apologize The data figures and descriptions contained in this man...

Page 2: ...2 Conductivity meter Thermometer HD3406 2...

Page 3: ...o Fahrenheit 9 Function indicators 10 key in the menu confirms the current selection 11 Line for symbols and comments 12 Main display line 13 ON OFF AUTO OFF key turns the instrument on and off When p...

Page 4: ...one or more of the 147 S 1413 S 12880 S or 111800 S cm conductivity standard solutions The displayed data can be stored datalogger and can be transferred to PC or serial printer thanks to the multi st...

Page 5: ...ment ready for normal measure ment MIN cm min s 2 3 m m k During turning on should no probes be connected the last manually set temperature ap pears in the secondary line The unit of measurement symbo...

Page 6: ...rks on command the sending of data to the serial port is performed each time the key is pressed Recording LOG is performed with one second intervals even if the interval is set to 0 With an interval f...

Page 7: ...you can see an example of the measurement process with the Auto Hold function enabled In the following figure you can see an example of the measurement process with the Auto Hold function enabled A pr...

Page 8: ...r to conduc tivity the indicated temperature is that measured at the maximum and minimum conductiv ity and is not the maximum and minimum temperature If the RCD MODE parameter is set to tp temperature...

Page 9: ...onductivity liquid resistivity total dissolved solids salinity and temperature meas urements updating them with the acquisition of new samples The acquisition frequency is once a second Use the TDS ke...

Page 10: ...e next the function is controlled by the Sleep_Mode_LOG parameter When the logging interval is less than one minute the logging instrument remains on with an interval of at least one minute it turns o...

Page 11: ...is the same for both the instrument and the PC the communication will only work in this way Erase memory When pressed after the MENU key the SERIAL key permanently erases all the data contained in th...

Page 12: ...12 Down Arrow When used in the menu it decreases the current variable value During measurement if the tem perature probe is not present it decreases the temperature value for conductivity compensation...

Page 13: ...on of the temperature sensors is not required The probes are detected during turn on and this cannot be performed when the instrument is already on therefore if a probe is connected and the instrument...

Page 14: ...ones for high values The indicative measurement range is reported in the following diagram Automatic or manual conductivity compensation The conductivity measurement refers to a standard temperature c...

Page 15: ...ution 147 S cm 25 C 0 01 Molar KCl solution 1413 S cm 25 C 0 1 Molar KCl solution 12880 S cm 25 C 1 Molar KCl solution 111800 S cm 25 C Using one of these solutions the calibration is automatic the pr...

Page 16: ...vity at the temperature at which the calibration is performed Proceed as follows 1 Turn the instrument on with the ON OFF key 2 Set the probe s cell constant by selecting it from the admitted values 0...

Page 17: ...a long beep The instrument remains in calibration mode and maintains the previous calibration values The most frequent causes of error are due to the probe malfunctioning deposits dirt or to the stand...

Page 18: ...at least 60mm the sensor is housed in the end part of the probe When measuring the temperature on frozen blocks it is convenient to use a mechanical tool to bore a cavity in which to insert the tip pr...

Page 19: ...ide The wires coming from the conductivity cell are welded directly onto the 1 2 4 5 connector s pins On the left there are the 1 4 points on which the Pt100 or Pt1000 sensor wires must be welded The...

Page 20: ...ns 1 and 4 Jumper between pins 2 and 5 4 electrode conductivity probe Internal view of the connector None As an alternative the Pt100 and Pt1000 sensors can be welded directly on the connector s pins...

Page 21: ...Jumper between pins 1 and 4 Jumper between pins 2 and 5 4 electrode conductivity probe Pt100 tempera ture sensor Pt100 Internal view of the connector None Ensure the welds are clean and perfect Once...

Page 22: ...ent is always very hard to perform It has high levels of uncertainty and depends on the ability of the operator 10 Temperature measurements on non metal surfaces usually require a great deal of time d...

Page 23: ...ou work with a black Platinum cell new platinum coating of the electrode could be necessary Cell constant value not accepted Check that the standard solutions are in good condition that the probe s co...

Page 24: ...full the instrument cannot store further data the memory space is full CAL blinking No calibration has yet been effected or the cell constant value has been modified in the menu please see the descri...

Page 25: ...OG logging in progress LOG MEM FULL memory full LOG_CLRD memory data cleared LOG_DUMP_OR_ERAS transfer or erase data LOG_STOP logging complete LOG_STRT logging started MIN USE_UNIT_TO_ZERO SEC minutes...

Page 26: ...ent issues a long beep and turns off In this case replace the batteries or use the external power supply in order to turn the instrument back on In order to avoid data loss the logging session is ende...

Page 27: ...to direct sunlight The instrument may be exposed to a source of high temperature The instrument may be exposed to strong vibrations The instrument may be exposed to steam salt or any corrosive gas MA...

Page 28: ...r where XY is the command code and cr is the Carriage Return ASCII 0D Command Response Description P0 Ping locks the instrument keyboard for 70 seconds P1 Unlocks the instrument keyboard S0 AT 21 3 10...

Page 29: ...seconds WC0 Setting SELF off WC1 Setting SELF on Command characters are exclusively upper case characters Once a correct command is entered the instrument responds with when any wrong combination of c...

Page 30: ...the LOG key the instrument logs the first data and turns off 15 seconds before the next logging instant it turns on again to capture the new sample and then turns off The data stored in the memory ca...

Page 31: ...t by pressing SERIAL again See the description of the menu items on page 6 Connect the S print BT printer using cable HD2110CSP NOTES The print out is formatted across 24 columns Some keys are disable...

Page 32: ...NU ENTER until the Baud Rate parameter select 38400 using the arrows confirm with ENTER The parameter remains in the memory until replacement of the batteries 4 Launch the DeltaLog9 application and pr...

Page 33: ...tion could take a few minutes 17 The installation procedure is now complete the device will be detected on each new connec tion automatically In order to check if the entire operation was successful i...

Page 34: ...egular functioning and operating security can be ensured only within the climatic conditions specified in the chapter TECHNICAL CHARACTERISTICS Do not use or store the instrument in ways and or places...

Page 35: ...sorbed with instrument off 20 A Mains cod SWD10 Output mains adapter 100 240Vac 12Vdc 1A Security of memorized data Unlimited Time Date and time Schedule in real time Accuracy 1min month max departure...

Page 36: ...Instrument Measurement range Kcell 0 01 Resolution Up to 1G cm Measurement range Kcell 0 1 Resolution Up to 100M cm Measurement range Kcell 1 Resolution 5 0 199 9 cm 0 1 cm 200 999 cm 1 cm 1 00k 19 9...

Page 37: ...0 200 C Pt1000 measurement range 50 200 C Resolution 0 1 C Accuracy 0 25 C Drift after 1 year 0 1 C year Automatic manual temperature compensation 0 100 C con T 0 00 4 00 C Reference temperature 20 C...

Page 38: ...K 0 01 0 04 S cm 20 S cm 0 120 C 2 electrode cell in AISI 316 Ultra pure water Measurement in closed cell SPT01G K 0 1 0 1 S cm 500 S cm 0 80 C 2 electrode cell in Platinum wire Probe material glass...

Page 39: ...4 wires 50 200 C Class A TP47 1000 Pt1000 2 wires 50 200 C Class A TP87 100 Pt100 4 wires 50 200 C Class A TP87 1000 Pt1000 2 wires 50 200 C Class A Temperature drift 20 C 0 005 C TP47 Module for the...

Page 40: ...nection of Pt100 4 wire and Pt1000 2 wire probes to instru ment series HD34 without amplifying electronics and linearization CONDUCTIVITY PROBES AND COMBINED CONDUCTIVITY AND TEMPERATURE PROBES SP06T...

Page 41: ...cable with connector length 2 metres TP87 100 Pt100 sensor immersion probe Probe s stem 3mm length 70mm Connection cable 4 wires with connector length 1 metre TP87 1000 Pt1000 sensor immersion probe P...

Page 42: ...mbined probes 4 wire Pt100 probes and 2 wire Pt1000 probes 18 WARNINGS AND OPERATING INSTRUCTIONS 22 Notes on conductivity measurement 22 INSTRUMENT SIGNALS AND FAULTS 24 LOW BATTERY WARNING AND BATTE...

Page 43: ...43...

Page 44: ...il pr sum d fectueux Dieser Garantieschein muss der Spedition beigelegt werden wenn das Ger t an das Kundendienstzent rum gesandt wird WICHTIG Die Garantie ist nur g ltig wenn dieser Abschnitt bis ins...

Reviews: