HD2301
- 15 -
V1.7
8.
Insert the gasket in the module;
9.
Screw the fairlead and the ring.
10.At the other end of the module, enter the ring with the O-Ring as indicated in the
picture.
11.Make sure the cable is not twisted while you are screwing the fairlead. Now the
probe is ready.
D
IRECT CONNECTION OF
4
WIRE
Pt100
SENSORS
4 wire Pt100 sensors can be soldered
directly to the pins of the flying female
connector without making use of the
TP47 board. The 4 wires of the Pt100
sensors have to be soldered as
indicated in the figure on the left. In
order to use this type of probe it is
necessary to set up the menu item
“Probe Type”, as described at page 9.
The P100 probe is recognized upon
turning on the instrument: connect the
probe when the instrument is switched
off and then turn it on.
Sensor
Direct soldering to the connector
Pt100 4
wires
View of the soldering side
of the flying female connector