Direct connection of 4 wire Pt100 sensors
4 wire Pt100
sensors
can
be soldered di-
rectly to the pins of the free female con-
nector without making use of the TP47
board. The 4 wires of the Pt100 sensors
have to be soldered as indicated in the
figure on the left.
In order to use this type of probe it is
necessary to set up the menu item “Probe
Type” as described at page 11.
The P100 probe is recognized upon turn-
ing on the instrument: connect the probe
when the instrument is switched off and
then turn it on.
Sensor
Direct soldering to the connector
4-wire Pt100
View of the soldering side
of the free female connector