background image

 

HD2024 

-   5   -

 V1.0

 

3

 

CALIBRATION PROCEDURE 

The HD2024 can calibrate standard 1/2” (and 1/4” with HD2020MA adapter) micro-
phones compliant with IEC 61094-1 and IEC61094-2. 
To calibrate the microphone, insert it deep into the cavity. The O-ring will offer some 

resistance. 
The calibrator can be held in vertical position or placed on a flat surface. 
While measuring, you should move neither the microphone nor the calibrator; make 

sure that the worktable doesn’t transmit vibrations. 
A small misalignment of the microphone and calibrator axes is allowed. 

 

Switch on the calibrator by pressing the ON/OFF key.  

 

Calibrate the sound level meter as per the procedure shown in the instrument 
manual. 

 

Apply the correction to the pressure level depending on the type of microphone 
(see the following paragraph). 

 

Once calibration is complete, switch off the sound level meter and the calibrator 
and remove the microphone from the cavity. 

The  HD2024 calibrator allows calibrating any sound level meter provided that it is 

equipped with a laboratory or working standard 1/2” microphone compliant with  IEC 
61094-1 and IEC 61094-4 standards. 

3.1

 

C

ORRECTION DEPENDING ON THE TYPE OF MICROPHONE

 

The  calibrator generates a 94 dB sound pressure level referred to 20 

μ

Pa. Working 

standard 1/2” microphones for sound level meters are manufactured to achieve flat 
frequency response in free or diffuse field, i.e. in a field of progressive plane waves 
propagating in the same direction as the microphone axis and in a field of sound 

waves coming from every direction, respectively. These propagation conditions are 
different from those in the calibrator cavity. 
In free field, reflections due to the microphone alter the sound level by increasing the 

actual high-frequency capsule sensitivity. Microphones optimized for free field meas-
urements exploit this phenomenon to achieve flat frequency response even at very 
high frequencies. In these microphones, the sound level increase at 1 kHz corre-
sponds to approximately 0.05 dB ÷ 0.20 dB. Therefore, when you calibrate a free field 

microphone, you must take into account for this  difference by setting in the sound 
level meter a sound level 0.1 dB or 0.2 dB lower than the calibrator nominal one. 
Microphones optimized for diffuse field measurements don’t require corrections when 

calibrated in a closed cavity at 1 kHz instead. 

 

Summary of Contents for HD2024

Page 1: ...Operating manual Acoustic calibrator HD2024 www deltaohm com English Keep for future reference Companies Brands of GHM...

Page 2: ...5 3 1 CORRECTION DEPENDING ON THE TYPE OF MICROPHONE 5 4 BATTERY 6 4 1 LOW BATTERY INDICATION 6 4 2 BATTERY REPLACEMENT 6 4 3 CAUTIONS ON USING THE BATTERY 6 5 CONSTRUCTION AND FUNCTIONING 7 6 TECHNI...

Page 3: ...94 4 Measurement microphones Part 4 Specifications for working standard microphones The generated sound pressure level is equal to 94 dB at 1000 Hz frequency A LED signals the low battery condition Ad...

Page 4: ...RIPTION 1 ON OFF key turns on and off the instrument 2 LED indicating the calibrator status 3 Calibrator cavity for 1 2 microphones 4 LOW BATT LED indicates that the battery is low and needs to be rep...

Page 5: ...with IEC 61094 1 and IEC 61094 4 standards 3 1 CORRECTION DEPENDING ON THE TYPE OF MICROPHONE The calibrator generates a 94 dB sound pressure level referred to 20 Pa Working standard 1 2 microphones...

Page 6: ...operation of the instrument the LOW BATT LED lights up 4 2 BATTERY REPLACEMENT To replace the 9 V battery Switch off the instrument Open the lid at the bottom of the instrument Replace the battery Cl...

Page 7: ...overpressure due to their insertion The electronics consist mainly of an oscillator combining high stability and low distor tion as well as an RMS converter the automatic gain control AGC the driver f...

Page 8: ...e Stabilization time 10 s Total distortion 1 Ambient condition influence Temperature and humidity influence Static pressure influence 0 3 dB in the range 10 50 C and 10 90 RH 0 1 dB in the range 65 10...

Page 9: ...operating safety can be ensured only if all stan dard safety measures as well as the specific measures described in this manual are followed The instrument proper operation and operating safety can be...

Page 10: ...struction manual HD2020MA Microphone adapter for 1 4 capsules DELTA OHM metrology laboratories LAT N 124 are ISO IEC 17025 accredited by ACCREDIA for Temperature Humidity Pressure Photometry Radiometr...

Page 11: ......

Page 12: ...zed third parties is excluded from the guarantee The instrument shall be returned FREE OF SHIPMENT CHARGES to your dealer The jurisdiction of Padua applies in any dispute The electrical and electronic...

Page 13: ...e result of the constant development of the product This may produce some differences between the information written in this manual and the instrument you have purchased We cannot completely exclude...

Page 14: ...GHM GROUP Delta OHM Delta Ohm S r l a socio unico Via Marconi 5 35030 Caselle di Selvazzano Padova ITALY Phone 39 049 8977150 Fax 39 049 635596 www deltaohm com info deltaohm com V1 0 04 06 2018...

Reviews: