HD2013
- 19 -
V1.5
Command
Description
PWD
CTRi
Reads the setting of the multiplication factor of index i.
Reply of the instrument: r…rk…k, with:
r…r = rainfall rate corresponding to the multiplication factor
k…k = value of the multiplication factor
Example: +50+1.05 indicates that the multiplication factor is 1.05
and the corresponding rainfall rate is 50 (mm/h or in/h).
No
When you set the correction table, it is mandatory to enter in ascending order all the
ten multiplication factors. The rate corresponding to the multiplication factor of index i
must be greater or equal to the rate corresponding to the multiplication factor of index
i-1.
Example
To implement the following correction table:
Rainfall rate (mm/h)
0…25 25…50 50…75 75…90 90…120 120…150 150…180 180…220 220…260 260...
K
1
1.02
1.04
1.06
1.08
1.1
1.12
1.15
1.18
1.2
send, exactly in the order shown, the following commands:
CTS
CTW0+0+1
CTW1+25+1.02
CTW2+50+1.04
CTW3+75+1.06
CTW4+90+1.08
CTW5+120+1.1
CTW6+150+1.12
CTW7+180+1.15
CTW8+220+1.18
CTW9+260+1.2
CTE
Note that the sequence of the CTW commands must be preceded by the command
CTS and follwed by the command CTE.