- 36 -
8) PSW_MENU
(password)
1)
RST_PSW_LVL
: exits the menu and deactivates immediately the password (the
password will not remain active for some minutes as it normally happens when ex-
iting the menu: you will need to re-enter the password even if you re-access imme-
diately the menu).
2)
SET_NEW_PSW
: sets user-level password.
3)
EXIT
: returns to the main menu.
9) CAL_MENU
(calibration) –
Only available with administrator password
1)
RH_75%_CAL
: relative humidity sensor calibration at 75%RH.
2)
RH_33%_CAL
: relative humidity sensor calibration at 33%RH.
3)
DIFF_PRES_0 Pa_CAL
: calibration of differential pressure to zero.
4)
LGHT_SENS_PA_LUX
: illuminance sensor sensitivity in pA/lux.
5)
UVA_SENS_nA_W/m
2
: UVA irradiance sensor sensitivity in nA/Wm
-2
.
6)
CAL_TYPE
: choice between user calibration (
USER
) or factory calibration (
FACT
).
7)
EXIT
: returns to the main menu.
Notes
:
- relative humidity calibration is available only with high accuracy R.H. sensor models:
HD35ED…TC and HD35ED…TV models.
- differential pressure zero calibration is available only in HD35ED…4r…TV models.
- the illuminance sensor sensitivity is available only in HD35ED…I…TCV models.
- the UVA irradiance sensor sensitivity is available only in HD35ED…IU…TCV models.
10) EXIT
Returns to measurement mode.