17
DL-00096 Rev. 7
Alarm Message
Check Vents
Low Battery
XX: Service!*
*Value: 21-50
Description
Device is unable to maintain
oxygen purity.
Estimated battery life less
than 17 minutes.
Service required.
Action
Be sure air inlet/outlet
has not been blocked. If
alarm persists, contact
Customer Service.
Charge the battery pack
by plugging in to power
supply.
Note: The message will be
automatically cleared when
plugged in to power supply.
Contact Customer
Service.
High Priority Alarms
Note: These alarms will allow the Portable Oxygen Concentrator to continue operating.
Alarm Message
Check Cannula
XX: Service*
*Value: 51-70
Description
No breath detected for 5
minutes.
Service required.
Action
Check the cannula con-
nection. Be sure to
breathe through nose. If
alarm persists, contact
Customer Service.
Note: The message will be
automatically cleared when
breathing is detected.
Contact Customer
Service.
Medium Priority Alarms
DL-00096 rev 7.indd 17
6/1/2009 4:48:34 PM