Using the 0 –100 Arbitrary Scale

This scale is used to test the moisture content of hygroscopic
materials for which a calibration is not available. Depending
on the material, a special application external electrode,
instead of the integral contact pins may be required.
Increasing readings on the 0-100 reference scale indicate
higher levels of moisture content. These readings can be
translated into percent moisture content once a calibration
has been developed.

h

Set the meter scale

for 

#

2. If necessary, attach an exter-

nal electrode to the meter.

h

Push the contact pins 

into the material or apply the exter-

nal electrode.

h

Press the read key 

!

. The meter displays a relative value

for two seconds.

The readings may also be used for comparative tests, after
meter readings have been related to given conditions for the
materials involved. When the meter is used as a gauge for
comparative tests, readings should be taken on samples con-
sidered to be at “safe” levels or in satisfactory condition.
These readings are then used as the “standard” against which
subsequent readings on the same material are evaluated.

The “standard” for any given material is related to safe stor-
ability or any other property which is important for further
production processing.

Testing Baled Scrap Paper

h

Set the meter scale for 

#

3 baled scrap paper.  Attach an

external electrode to the meter.

h

Push the external electrode into the material being tested.

h

Press the read key

!

. The meter displays the %MC for

two seconds.

The level of accuracy of meter readings depends on a number
of factors: similarity between the material tested and sam-
ples on which the calibration was made; moisture distribu-
tion; and chemical application or processing which may
affect the electrical properties of the paper product

The required electrode is the H-4 with a 

#

830-series prod.

(10”or 18”). A sharp, steel rod to open the hole for the prod
may be helpful if the bale is very dense.

7

www.

GlobalTestSupply

.com

Find Quality Products Online at:

[email protected]

Summary of Contents for P-2000

Page 1: ...P 2000 owner s manual W H E N A C C U R A C Y I S T H E P O I N T TM www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Page 2: ...oint 5 Information About Your Readings 5 To Check the Accumulated Readings 6 To Reset the Meter 6 Applications 7 Using the 0 100 Arbitrary Scale 7 Testing Baled Scrap Paper 8 Care of Your Meter 9 Service for Your Meter 10 Warranty www GlobalTestSupply com Find Quality Products Online at sales GlobalTestSupply com ...

Page 3: ... microcontroller circuit h Averages up to 100 accumulated readings h 4 3 18 moisture range on paper h 0 100 reference scale for relative moisture indications in hygroscopic materials h 5 40 moisture range on baled scrap paper h Includes 1 9 volt battery h Includes sturdy carrying case h One year warranty Read Key Calibration Check Key Set Point Key Scale Key 2 D E L M H O R S T P 2 0 0 0 www Globa...

Page 4: ...ng calibration there is no need to disconnect the external electrode if attached If you check the calibration and the display does not read 11 1 it is likely an indication of a low battery If this occurs change the battery immediately Continued use with a low battery may cause the meter to go out of calibration If you have a fresh battery and the instru ment still does not indicate an acceptable c...

Page 5: ...current set point is displayed and scroll to the set point value desired hTo scroll backward through the set point values press and release the set point key Within one second press and hold the scale key Continue to hold the scale key and the set point will decrease hWhen scrolling in either direction release the key to stop at your desired set point hA buzzer sounds if the meter reads a value hi...

Page 6: ...ic material be sure to clear the meter before moving on to the next scale if you do not want to group all of the readings together TO CHECK THE ACCUMULATED READINGS This feature displays the total number of all accumulated read ings for the selected material the average of those readings and the highest stored reading hTo view the readings press and release the calibration check key First the mete...

Page 7: ...ecked on the specific material and corrections determined by the user Meter readings indicate moisture content at room tempera ture of 70º F 90ºF Meter readings will be affected by lower or higher temperatures Lower temperatures cause readings to be lower higher temperatures cause readings to be higher than the actual MC The meter tends to read the highest moisture content that is in contact with ...

Page 8: ... be taken on samples con sidered to be at safe levels or in satisfactory condition These readings are then used as the standard against which subsequent readings on the same material are evaluated The standard for any given material is related to safe stor ability or any other property which is important for further production processing Testing Baled Scrap Paper hSet the meter scale for 3 baled s...

Page 9: ...terpret the readings hReadings of 5 10 with EMC to 60 RH are usually considered dry hReadings from 11 20 with EMC to 95 are usually considered acceptable but should be taken with some reservation hReadings of 20 40 are considered wet and unac ceptable CARE OF YOUR METER To keep your meter in good working order hStore your meter in a clean dry place The protective carrying case provided is an ideal...

Reviews: