background image

 
 

ENHANCED MODE 

 

1.  With meter powered ON, scroll to #6 (JOB DEF) and press SELECT. 
2.  Using  

 key scroll to assigned job (1 to 5). 

3.  Using 

key scroll to DONE and press SELECT. 

4.  To assign job type, press SELECT. Then press 



 to toggle between 

COMMERCIAL and RESIDENTIAL.  Scroll to DONE and press SELECT. 

5.  To assign ID press SELECT. Use 



 to change ID number. Scroll to 

DONE and press SELECT. 

6.  Assign Material Temp (70°F if unknown, only required on wood). 
7.  Assign location definitions by highlighting LOC DEF and pressing 

SELECT. 

a.  Using 

 

scroll to rooms to be tested, using 



 to assign rooms. 

b.  For each room assigned, press SELECT to change materials for 

that room. 

c.  When done assigning rooms, scroll to DONE and press SELECT. 

8.  Scroll to DONE and press SELECT. 
9.  Choose either SCAN MODE or PIN MODE and press SELECT. 
10.  Take reading as in earlier instructions. 
11.  Press SELECT to store the reading.  The meter will automatically 

advance to the next assigned location, as shown in lower left hand corner. 

12. Repeat readings until all locations have been stored. 

 

 

THERMO-HYGROMETER MODE   

 

1.  Insert RH/Temp sensor to connector on top of meter. 
2.  Using 



 

scroll to RH-T MODE. 

3.  Press SELECT 

NOTE:    You  may  take  temperature/RH  readings  in  both  BASIC  and 
ENHANCED  modes.    In  ENHANCED  mode,  you  must  press  SELECT  to 
hold the reading and then press it again to move to the next location. 

 

 

IR LINK 

 

1.  You must have Microsoft Excel to use the IR Link Software. 
2.  Install the software and IR Link as per the instruction manual. 
3.  Place the Navigator with the Infrared eye pointed toward the eye on the IR 

Link. 

4.  Using buttons scroll to #7 (IR Link) and press SELECT. 
5.  Do not touch until linking is completed.

 

 
 
 

 

 

510INS-0027/QR 

 

 

01/14 

Reviews: