OXFAM-DELAGUA Water Testing Kit — Users Manual
63
Temperature calibration
The procedure for checking and calibrating the temperature of the 2 incubators is
the same as described in Section 7.3 for the single incubator kit. However, the
corresponding temperature adjuster screws are located underneath the labels
showing the incubator temperatures.
The temperature adjuster screws are located underneath the temperature labels
+
+