7 User Interface
26
Figure 17
he function bar provides the following options:
Program Start
Starts the program. opens the
Program Procedure
.
T
Program
Start
Å
Position
Æ
Position
Program
Display
Tc 550°C
# 123456789
00:12:22
011 Wash
firing
°C
980°C
550°C
min
Drying
04:00
Å
Æ
Browse through program procedure.
Program Display
Displays program entries.
Setup
Opens the
Setup
menu.
If an empty memory was selected when selecting the program number, the message
Program Edit
Create a new program. Opens the
Program Editor
menu
.
"This memory contains no data" is displayed instead of the entry boxes (see Figure 11).
Fig. 11). You now have the following options (function bar):
ProgramManager
Opens the
Program Manager
menu.
Setup
Opens the
Setup
menu.
7.4 PROGRAM PROCEDURE
A firing program is started from the
Program Display
by pressing Program Start .
TIP E
TIP A
s e e p . 4
s e e p . 4
This can also be done from the graphic display of the program procedure. The program
steps are processed according to the set sequence of the entry mask (Standard or
Professional, or press, (only the D4
press
) cf. Section 7.5). An exception is the Vacuum
program step. After closing, the furnace is evacuated to [VAC level] or [Press level] for
the D4
press
. At the temperature [VAC off] the furnace is vented after expiration of the
holding time [VAC hold], or after [Press time] for the D4
press
. A program procedure as
in Fig. 18 executes the following firing process: