Rooster
™
Sensor100 USER MANUAL
Rooster
TM
Sensor100 USER MANUAL 62320MN000-A04
19
2.
Push the “Airflow Settings” button to enter the Airflow
Settings menu screen.
Home>Settings>System>Airflow Settings
3.
In order to
proceed, you will need to enter your “Certify” password. If you have not set a
custom password, the factory default is
“6425”
4.
Push the “Volumetric Flow Settings” button to enter the Volumetric Flow menu screen.
Home>Settings>System>Airflow Settings>Volumetric Flow Settings
5.
Push the “Volumetric Flow” button until it says Enabled
.
Home>Settings>System>Airflow Settings>Volumetric Flow Settings>Volumetric Flow
6.
Push the “Duct Size” button to enter the Duct Size screen
.
Home>Settings>System>Airflow Settings>Volumetric Flow Settings>Volumetric Flow>Duct Size
7.
Press the “Shape” button until the appropriate duct shape is displayed, circular or
rectangular.
8.
Press the “Un
its” button to display your choice of cm or inch units for entry of duct
dimension(s).
9.
Enter the dimension(s) of the duct, diameter for circular shape, height, and width for
rectangular shape.
A third units touch toggle of cfm is accessible on the main screen with volumetric flow enabled
so that the user is able to cycle between fpm, m/s, and cfm.
Analog Output
The Rooster
TM
Sensor100 has digital to analogue converter (DAC) output enabled, with
correspondence to air velocity. There are four user selectable ranges to choose from, which the
user will connect to pins 7
& 8 on the rear panel’s 8
-
position “Outputs” Connector:
o
0-5 VDC