Management Procedure 2559 Rev A 

 

Page 

2

 of 

7

 

1

 

Introduction and UUC Performance Requirements 

 

1.1

 

This procedure describes the calibration of DeFelsko Corporation PosiTest LPD.  

Table 1-1  

Models 

Measurement Range 

PosiTest LPD 

80 – 400 k

Ω

  

9 – 90 VDC 

 
1.2

 

The unit being calibrated will be referred to as the UUC (unit-under-calibration). 

    

2

 

Measurement Standards and Support Equipment Performance Requirements 

 
2.1

 

The UUC accuracy requirements are based upon the published UUC performance 
specifications. 

 
2.2

 

The test uncertainty ratio applied in this Calibration Procedure is 4:1 unless otherwise 
stated. 

 
2.3

 

The Minimum-Use-Specifications are the minimum test equipment specifications 
required to meet all the UUC accuracy requirements and the test uncertainty ratio 
applied. 

Table 2-1 UUC Accuracy Requirements and Description 

UUC Parameter 

Performance 

Specifications 

Test Method 

Resistance 

80 – 400 k

Ω

 + 3% 

Resistance Decade 

Box 

DC Voltage* 

9 – 90 V + 5%  

Digital Multimeter 

* Optional calibration 

 

Table 2-2 Minimum Use Specification 

Parameter 

Range 

Accuracy 

Resistance 

80 – 400 k

Ω

 

0.6 k

Ω

 

DC Voltage 

9 – 90 V 

0.1 V 

 

Table 2-3 Actual Equipment Specification 

Parameter 

Range 

Accuracy 

Manufacturer/Model #’s 

Applicable 

Resistance 

0 – 9,999 k

Ω

 

+(0.1% of r 0.025

Ω

IET Labs 

RS-201 

DC Voltage 

5 – 50 V 

+(0.02% of r 0.004V) 

Keysight 

U3401A 

DC Voltage 

51 – 500 V 

+(0.02% of r 0.04V) 

Keysight 

U3401A 

 

Summary of Contents for PosiTest LPD

Page 1: ...ments 2 Table 1 1 2 2 Measurement Standards and Support Equipment Performance Requirements 2 Table 2 1 UUC Accuracy Requirements and Description 2 Table 2 2 Minimum Use Specification 2 Table 2 3 Actual Equipment Specification 2 Table 2 4 Calibration Environmental and Warm up Requirements 3 3 Resistance Calibration 3 4 DC Voltage Calibration 5 5 Performance Requirements 6 Table 5 1 Performance Requ...

Page 2: ...rwise stated 2 3 The Minimum Use Specifications are the minimum test equipment specifications required to meet all the UUC accuracy requirements and the test uncertainty ratio applied Table 2 1 UUC Accuracy Requirements and Description UUC Parameter Performance Specifications Test Method Resistance 80 400 kΩ 3 Resistance Decade Box DC Voltage 9 90 V 5 Digital Multimeter Optional calibration Table ...

Page 3: ...tion process 3 1 Review the Performance Requirements in Table 5 1 Note Whenever the test requirement is not met verify the results of each test and take corrective action before proceeding 3 2 Turn on the UUC by pressing the power button The unit will complete a self test indicated by flashing each of the 4 voltage lights Once the self test is completed the LED corresponding to the set voltage wil...

Page 4: ...the UUC stops alarming or alarms intermittently Record the last resistance value that the unit alarmed constantly on Record this value as kΩ For example 89 kΩ and 700 Ω would be recorded as 89 7 kΩ 3 4 2 If the UUC is not alarming or is alarming intermittently decrease the resistance value until the UUC starts alarming constantly Record this resistance value 3 4 3 Set the UUC to 67 5V 80 kΩ and ad...

Page 5: ...e 5 1 Note Whenever the test requirement is not met verify the results of each test and take corrective action before proceeding 4 1 1 Set up the UUC as described in sections 3 2 and 3 3 4 1 2 Turn on the multimeter and set it to DC Volts and auto ranging 4 1 3 Connect the positive and negative leads from the multimeter to the corresponding terminals of the RS 201 decade box 4 1 4 Set the UUC to 9...

Page 6: ...nts and Calibration Data for PosiTest LPD Parameter Nominal Min1 UUC Reading Max2 Resistance 90 kΩ 9V 87 3 92 7 Resistance 80 kΩ 67 5V 77 6 82 4 Resistance 90 kΩ 67 5V 87 3 92 7 Resistance 400 kΩ 90V 388 412 DC Voltage 9V 90 kΩ 8 55 9 45 DC Voltage 67 5V 80 kΩ 64 13 70 87 DC Voltage 67 5V 90 kΩ 64 13 70 87 DC Voltage 90V 400 kΩ 85 50 94 50 1 Resistance min nominal value kΩ 0 97 DC Voltage min nomi...

Page 7: ...evel A Date of Change November 9 2016 Title Calibration Procedure PosiTest LPD Low voltage Pinhole Detector Reason for Change New product Description of Change New procedure I confirm I have read and understand the procedure and the change described above Printed Name Signature Date Management Form 0010 02 05 1998 ...

Reviews: