Management Procedure 2559 Rev A
Page
1
of
7
Management Procedure 2559
Revision: A
Date Issued: November 9, 2016
Date Revised:
Calibration Procedure
PosiTest LPD Low voltage Pinhole Detector
Table of Contents
1
Introduction and UUC Performance Requirements ................................................................. 2
Table 1-1 .................................................................................................................................. 2
2
Measurement Standards and Support Equipment Performance Requirements ........................ 2
Table 2-1 UUC Accuracy Requirements and Description ....................................................... 2
Table 2-2 Minimum Use Specification .................................................................................... 2
Table 2-3 Actual Equipment Specification .............................................................................. 2
Table 2-4 Calibration Environmental and Warm-up Requirements ........................................ 3
3
Resistance Calibration .............................................................................................................. 3
4
DC Voltage Calibration* .......................................................................................................... 5
5
Performance Requirements ...................................................................................................... 6
Table 5-1 Performance Requirements and Calibration Data for PosiTest LPD ...................... 6