3
Carousel:
indicates the current view. Swipe horizontally
or use arrow keys to switch between views.
Pull Parameters
(pg. 10)
WiFi Bluetooth USB
(Status icons)
Pull Chart
displays
pressure
over time
Current
Reading
Touch Screen Display
Dolly
Size
Pull Rate
Limit Hold Time
Use the touch screen to interact with menu options. Swipe
horizontally
to navigate between views or
vertically
to move
between batch readings. Alternatively, the directional buttons can
be used for navigation.
Battery
(pg. 7)
Main View
(shown with memory batch open)
Dolly & Coating Preparation
Dolly Preparation
1.
To remove oxidation and contaminants, place the included
abrasive pad on a flat surface and rub the base of the dolly
across the pad 4-5 times.
2.
As required, remove residue left from the abrading process
using a dry cloth or paper towel.
Coating Preparation
1.
Lightly roughen the coating using the included abrasive pad.
As coating abrasion may introduce flaws, it should
only be used when necessary to remove surface contaminants,
or when the bond strength between the glue and the coating
is insufficient for pull testing.
NOTE: