Page 5
Setup
Set configuration options for the Gage.
Set Clock
All measurements are date and time stamped (24 hour format) when stored into
memory. It is therefore important to keep both the date and time current using this menu
option. Use the Up and Down buttons to scroll, and the (-) and (+) buttons to adjust
value. The Gage’s date and time can also be viewed in Gage Info and on top of the
main menu.
Reset
Restores factory settings and returns the Gage to a known condition. The following
occurs:
All batches, stored datasets, batch names and screen captures are erased.
Menu settings are returned to the following:
Memory = OFF
Auto Log = OFF
Trend Chart = None
Auto Dim = ON
Display = None
Bluetooth & Stream = OFF
WiFi & Access Point = OFF
USB Keyboard & Stream = OFF
Bluetooth BLE Keyboard = OFF
Anemometer = ON (DPM-A only)
Perform a more thorough Hard Reset by powering down the Gage, waiting several
seconds, then simultaneously holding both the center navigation and (+) buttons until
the Reset symbol appears. This returns the instrument to a known, out-of-the-box
condition. It performs the same function as a menu Reset with the addition of:
Bluetooth Pairing info is cleared. Alarm settings are set to defaults. Menu settings are
returned to the following status:
Units (temperature) = Celsius