4
Standard models
will alert the user when the surface temperature
is less than 3°C (5°F) above the dew point temperature. Enable by
selecting the tick box.
The
PosiTector
DPM
can automatically alert the user when
current climatic conditions exceed pre-set values. When selected,
the Alarm icon is displayed at the the top of the screen.
Advanced models
allow the user to set up custom alarm
conditions for each reading in a dataset. When connected to WiFi,
Advanced models can also send alerts via email.
Alarm Mode
Alarm Mode
Alarm
Setup Menu
Setup Menu
Units
Converts the displayed units for temperature and wind velocity.
Temperature
(all models)
: Celsius (°C) or Fahrenheit (°F)
Velocity
(DPM A only)
: m/s, ft/s, km/h, mph, or kn.
Reset
Reset
(menu reset) restores factory settings and returns the Gage
to a known condition. The following occurs:
- All batches and stored datasets, batch names and screen
captures are erased.
- Menu settings are returned to the following:
Memory
= OFF
Auto Log
= OFF
Trend Chart
= None
Auto Dim
= ON
Display
= None
Bluetooth & Stream
= OFF
WiFi & Access Point
= OFF
USB Keyboard & Stream
= OFF
BLE Keyboard
= OFF
Anemometer
= ON
Perform a more thorough
Hard Reset
as follows:
1.
Power down the Gage and wait 5 seconds.
2.
Simultaneously press and hold the
and buttons
until the
Reset
symbol appears.
This returns the Gage to a known, “out-of-the-box” condition.
www.
GlobalTestSupply
.com
Find Quality Products Online at: