Technical Specifications and Operating Procedure. PILATUS3 S 6M-NSRRC Detector System. Version: V1.0
28/33
8.
Turning off the Detector
In-air operation:
Turn OFF the power switch at the back of the detector electronics.
Do not remove the nitrogen/dry air connection and leave it at the recommended flow rate
according to section 5.6.
In-vacuum operation:
Turn OFF the power switch at the back of the detector electronics.
Keep the detector under vacuum.
In-vacuum operation - prior to venting (and pumping down):
Always make sure the detector is SWITCHED OFF and WARMED UP to room temperature.
Otherwise it could be damaged through electrical discharge or condensation.
If you turn off the detector while Camserver is running you will get error messages after a few
minutes because Camserver cannot communicate with the detector. You may want to exit Camserver.
You must restart Camserver after the detector has been turned on! Otherwise the detector is not
initialized.