Figure 5.2:
Drawing of the PILATUS3 R CdTe 100K-M Detector Head with extended Mounting Brackets (top view)
Mounting the Detector Head from Below
Warning
#5
It is strictly forbidden to add any threads to the detector base plate or to the detector housing.
The detector should be mounted using the two internal M6x1 threads as shown in figure 5.3 (indicated with yellow circles).
Figure 5.3:
Drawing of the PILATUS3 R CdTe 100K-M detector head base plate (a printed copy is supplied in the user documentation
folder). Left: View from below with retracted mounting brackets for mounting from below. Right: Cross section view through the internal
thread to show the 7 mm thread-less and the 9 mm threaded sections.
PILATUS3 R CdTe 100K-M Technical Specifications v1.0.0
19 | 29