Decagon AquaLab Quick Start Manual Download Page 6

V S A   Q u i c k   S t a r t   G u i d e

2. AquaLab VSA Performance Verification overview—

  •  AquaLab VSA performance and cleanliness should be

 

    checked frequently; before each isotherm test is 

 

    recommended.

 

  •  Press   until you are at the Measurement screen (2nd tab).

  •  Use AquaLab Verification Standards to verify performance

 

    or adjust linear offset.

  •  To verify, remove weigh pan and then place sample cup

 

    containing one of AquaLab’s verification standards in

 

    sample chamber. Close lid and carefully avoid spillage; See

 

    Figure 5: Sample Chamber.

  •  Move latch left and/or press   to start a reading. The

 

    pinwheel will spin indicating a measurement is running.

 

    The 

a

w

 reading for the standard and the temperature

 

    will display when finished. Make sure the 

a

w

 reading

 

    on the standard is /-0.005 of the correct 

a

w

 for the

 

    standard. (+/- 0.02 if cap sensor is used)

  •  If an offset adjustment is needed, refer to chapter 7 in the

 

    user’s manual.      

Figu

re 5: Sample chamber

Lid

Base

V S A   Q u i c k   S t a r t   G u i d e

3. Sample Preparation—

  •  Fill sample cup no more than half full (~7ml). 

 

   

Do Not Overfill!

  •  Minimum sample amount should cover bottom of sample

 

    cup (500-2000 mg) (small gaps OK).

  •  Ensure rim and outside of sample cup are clean.

  •  Make sure sample temperature is not more than 4°C above

 

    chamber temperature.

4. Starting a Test—

  •  Before starting a test, set the Volatiles: Yes/No option in

 

    the configuration menu. If your sample has volatiles,

 

    select “Yes” otherwise select “No”.

  •  On the main screen, select the test wizard  .

  •  Provide a name for the test using the arrow keys and then

 

    press  . (The name of a previous test will appear by

 

    default.)

  •  Enter initial moisture content if known (optional). Press 

 

    to continue or   to go back.

  •  Setup each stage of the isotherm test. A stage is used to

 

    adjust the isotherm settings for an individual sample (up to

 

    20 stages). Each stage is setup with the isotherm method 

 

    (DDI or DVS) and the settings for that portion of the

 

    isotherm. Once a stage is created, it can be modified or

 

    deleted anytime before the test is started. 

 

 

   

i.e. For milk powder, Stage 1 could use the DDI method to  

    ramp quickly from 0.1 to 0.4 

a

w

 and Stage 2 could use 

    the DVS method to step from 0.4 to 0.8 

a

w

 slowly in 

    0.05 

a

w

 increments.

  

  •  To setup a new stage, select “new” and press  . A screen

 

    with several parameters will appear. If a previous test

 

    has been run, the stages for that test will be listed by

 

    default. These stages can be used as is or updated with

 

    new settings. If you need to delete a stage, go to the

 

    specific stage and press  . Scroll down to the bottom of

 

Reviews: