STEP 4
Apply Control Solution.
Move your finger (with control solution on
it) to meet the absorbent hole of the test
strip and the drop will be automatically drawn
into the test strip. Remove your finger until
the confirmation window is filled. The meter
begins to count down.
To avoid contaminating the control solution
with the content of the test strip, you have
to place a drop of control solution on a clean
surface or on your finger tip first. Then touch
the test strip to the drop.
STEP 5
Read and Compare the
Result
After counting to 0, the test result of control
solution is shown on the screen. Compare this
result with the range printed on the test strip
vial. It should fall within this range.
Out-of-range results
If test results fall outside the range printed on the test strip vial, check the section
of “
Problem in Operation
” in troubleshooting guide and repeat the test. If you
continue to get out-of-range results, it means that the system may not be working
properly. Do NOT test your blood. Contact your local Customer Care Line at
1(866) 829-7926 for help.