27
26
4.3
Internal code check/display
Click[ ← ] Shift, [ ↑ ] Increment, modify to the calculated
scale parameter
Modify BIT0 in SPEC 05 to "1"
Remove the heavy object, and press [REZERO] to enter the
zero confirmation and wait, click [REZERO] again to enter the
zero confirmation
Put on the weight, and display the actual weight value of the
modified weight
Remove heavy objects and return to weighing mode
Note:
1. When there are items on the load-bearing table or zero confirmation is
not required, you can press the [ON/OFF] key to directly exit the calibration
mode and return to the weighing mode.
2. The calibration switch (SPAN SW) must be turned on.
3. Calculation of the new standard rate:
Actual weight W1=10.00kg Display weight W2=10.005kg The original
standard rate of the instrument is V1= 37.500
Then the correction rate V2 = W1*V1/W2=10.00*37.500/10.005=37.48
Click [REZERO] button to save
Click the switch below the scale body to open the calibration
Long press [REZERO] key, click [ ↑ ] [ ↑ ] [ ← ] to enter the
internal code check
Click [ ↑ ] button to display [A/D] internal code
Click [ON/OFF] key to return to weighing mode
REZERO
ON
OFF
4. CALIBRATION OPERATION
4.2 Full scale rate correction
4.3 Internal code check/display
4. CALIBRATION OPERATION